Accuracy and Precision of Measured Blood Sugar Values by Three Glucometers Compared to the Standard Technique

Journal Title: Journal of Clinical and Diagnostic Research - Year 2017, Vol 11, Issue 4

Abstract

Introduction: The prevalence of diabetes mellitus has magnified, globally. Being cost effective and user friendly, using the glucometer systems has become more widespread. However, despite their advantages, there are controversies regarding their accuracy and precision. Aim: To determine the accuracy and precision of the common glucometers of Bionime, Accu-check, and Beurer compared with the laboratory technique. Materials and Methods: In a cross-sectional study, 257 patients of a general internal unit of Imam Reza hospital of Kermanshah were recruited. The blood sample of the subjects was transferred to the laboratory, then, two samples of capillary blood were tested by the three glucometers. The data was entered into a checklist, then entered to SPSS-16.0 software, and analyzed by descriptive and inferential statistics. Results: In this study, there was a positive correlation between the three blood glucose measurement methods with the standard approach, significantly, which indicates acceptable accuracy. However, this correlation was higher in the Beurer Glucometer (r=0.976, p<0.001). There were no significant differences in repeated measurements of blood glucose in the glucometers of Beurer (t=0.96, p=0.92), and Bionime (t=0.50, p=0.61) by paired t-test, but this test was significant for Accu-chek system (t=2.7, p=0.006). Conclusion: With respect to the acceptable accuracy and precision of Beurer and Bionime glucometers, it is suggested to use these methods for diabetic patients at homes as well as in the hospital units.

Authors and Affiliations

Shadi Khakpour Kermani, Alireza Khatony, Rostam Jalali, Mansur Rezaei

Keywords

Related Articles

The Role and Scope of Research in Undergraduate Medical Curriculum

ABSTRACT A continual enhancement of research capacity remains the cornerstone of modern medicine. Medical students need to understand, learn and apply the principles of research to practice of medicine. The best time to...

Radioprotective Effect of Beta D-Glucan and Vitamin E on Gamma Irradiated Mouse

Introduction: It is shown that beta-D-glucan is an immunologic system booster with radioprotectory effects. Radioprotectors are chemical components that can alleviate biological damage produced by ionizing radiation. Aim...

An Analysis of Dynamic Pulmonary Functions of Hypothyroid Patients

Introduction: Hypothyroidism is a silent epidemic of our times. In India, the prevalence of hypothyroidism is 11%. The effect of hypothyroidism on respiratory system is debatable with studies suggesting both obstructive...

Ankylosing Spondylitis Causing Obstructive Lung Disease and Tricuspid Insufficiency: A Case Report

ABSTRACT Ankylosing Spondylitis (AS) is a chronic progressive inflammatory spondyloarthropathy typically involving the axial skeleton, especially the sacroiliac joints and spine. Genetic predisposition in AS leads to its...

Photographic Assessment of Cephalometric Measurements in Skeletal Class II Cases: A Comparative Study

Introduction: Cephalometry has many limitations of which radiation exposure is most important. Hence, there is a need to resort to other safer methods which could give equal if not better results. Aim: The purpose of thi...

Download PDF file
  • EP ID EP337880
  • DOI 10.7860/JCDR/2017/23926.9613
  • Views 63
  • Downloads 0

How To Cite

Shadi Khakpour Kermani, Alireza Khatony, Rostam Jalali, Mansur Rezaei (2017). Accuracy and Precision of Measured Blood Sugar Values by Three Glucometers Compared to the Standard Technique. Journal of Clinical and Diagnostic Research, 11(4), 5-8. https://europub.co.uk/articles/-A-337880