Accuracy of densitometry of two cone beam computed tomography equipment in comparison with computed tomography

Journal Title: Electronic Physician - Year 2017, Vol 9, Issue 5

Abstract

BACKGROUND: The use of oral implants has been growing, and cone beam computerized tomography (CBCT) has become the method of choice for oral and maxillofacial radiology. OBJECTIVE: To assess the accuracy of bone densitometry in two different CBCT devices in comparison with MDCT (multi-detector CT). METHODS: Different concentrations of urografin, including 2.5%, 5%, 7.5%, 10%, 12.5%, were prepared, and the Hounsfield unit of these solution was measured by two CBCT devices (SORDEX CRANEX 3D and NEWTOM 5G) and one spiral CT device (SOMATOM SENSATION). Difference of output Hounsfield units in each concentration was compared in three devices. Correlation of devices with increase of urografin dose also was evaluated. Statistical analyses of the data were performed using SPSS18 and Kruskal-Wallis and Mann-Whitney U tests, along with Spearman's correlation coefficient. RESULTS: The range of gray density for NEWTOM 5G CBCT, SORDEX 3D CBCT, and SOMATOM CT imaging systems was from 781 to 2311, 427 to 1464, and 222 to 994, respectively. There was significant difference between devices in the Hounsfield unit in all urografin concentrations (p<0.001). Also there was a significant correlation between three devices with increasing the urografin dose (p<0.05; r>0.95). CONCLUSION: Our findings indicated a high correlation and linear relationship between different studied imaging systems. Although utilizing CBCT in the assessment of bone density is useful according to its lower emitted dose and less cost, clinicians should be aware of the issue that the voxel value in CBCT is not as perfect as CT

Authors and Affiliations

Sina Haghanifar, Sajad Yousefi, Ehsan Moudi, Farida Abesi, Ali Bijani, Ali Akbar Moghadamnia, Mehrdad Nabahati

Keywords

Related Articles

Combined trabeculotomy-trabeculectomy versus Ahmed valve implantation for refractory primary congenital glaucoma in Egyptian patients: a long-term follow-up.

Primary congenital glaucoma (PCG) is the most common type of glaucoma in pediatric patients. The aim of this study was to compare the effectiveness of combined trabeculotomy-trabeculectomy versus Ahmed valve implantation...

Comparing the quality of preconception care provided in healthcare centers in Mashhad in 2012.

Improving the quality of healthcare services is considered as the main strategy to improve maternal and neonatal health outcomes. Providing appropriate healthcare for mothers and their newborn children is facilitated sig...

Effect of Attachment Styles to Parents on Sexual Dysfunction Domains of Married Women

INTRODUCTION: According to Bowbly attachment theory, attachment of a baby and its main care provider, influences on social growth and the baby's feelings throughout its life. The present study was performed aim to determ...

Assessment of refractive outcome of femtosecond-assisted LASIK for hyperopia correction

INTRODUCTION: Laser vision correction for hyperopia is challenging. The purpose of the study was to assess the refractive outcomes of femtosecond-assisted laser in situ keratomileusis (LASIK) for hyperopic correction usi...

The impact of educational intervention on physical activity, nutrition and laboratory parameters in type II diabetic patients

BACKGROUND: Diabetes is one of the most common chronic diseases that is now widely spread around the world, and day by day the number of patients is increasing. The purpose of this study was to determine the impact of ed...

Download PDF file
  • EP ID EP321197
  • DOI 10.19082/4384
  • Views 101
  • Downloads 0

How To Cite

Sina Haghanifar, Sajad Yousefi, Ehsan Moudi, Farida Abesi, Ali Bijani, Ali Akbar Moghadamnia, Mehrdad Nabahati (2017). Accuracy of densitometry of two cone beam computed tomography equipment in comparison with computed tomography. Electronic Physician, 9(5), 4384-4390. https://europub.co.uk/articles/-A-321197