Agreement of Axial Length Measurements by Three Different Biometric Devices

Journal Title: Journal of Clinical and Diagnostic Research - Year 2018, Vol 12, Issue 6

Abstract

ABSTRACT Introduction: Accurate Intraocular Lens (IOL) power calculation is crucial for spectacle independence and patient satisfaction post cataract surgery. Of the various parameters measured to calculate IOL power, Axial Length (AL) is the most crucial. There have been several advances in AL measurement devices with the current gold standard being Optical Biometry (OB). However, OB cannot be used in all cases and hence the earlier Ultrasound (US) biometry continues to be relevant. It is however important to conduct agreement analysis with the gold standard to ensure accuracy of results in all cases. Aim: To compare the AL measurements obtained by Tomey AL 4000 US biometer and pachymeter (Tomey US), built in US biometer of the Nidek AL Scan biometer (Nidek AL US), and the Nidek AL partial coherence interferometry optical biometer (Nidek AL OB). Materials and Methods: This was a non-interventional cross-sectional observational study. Patients with immature cataract satisfying the inclusion criteria scheduled for surgery and undergoing biometry underwent AL measurements by the three methods. AL measurements obtained with the two US devices were compared with OB measurements and agreement analysis was done. Results: A total of 98 eyes of 60 patients were studied. The average AL measured by the three devices were, 22.97±0.79 mm by Tomey US, 23.24±0.89 mm by the built in applanation US of Nidek AL US scan and 23.08±0.81 mm by the Nidek AL OB. The mean inter-device difference in AL between Tomey US and Nidek AL OB was -0.11±0.15 mm and 0.16±0.44 mm between Nidek US and Nidek OB. There was excellent correlation between the Nidek AL OB and the Tomey US, Interclass Correlation was (ICC)=0.98, 95% CI (0.95, 0.99). Bland Altman Plot analysis also showed high agreement between these two devices. About 94% of the values were within 0.3 mm difference between these two devices and there were no eyes with ≥1 mm difference. Conclusion: There was excellent correlation of AL as measured by Nidek AL OB and Tomey US.

Authors and Affiliations

Anika Amritanand, Arathi Ro ddam Simha, Ringsar Baglari, Jenifer Roselin, John Michael, Gowri Mahasampath

Keywords

Related Articles

Intramuscular Capillary Haemangioma of the Temporalis Muscle: A Rare Case with A Review of the Literature

An Intramuscular Haemangioma (IMH) is a benign mesenchymal tumour of the endothelial cells that accounts for less than 1% of all haemangiomas. Here we report the case of a capillary type intramuscular haemangioma in a fi...

A Stratified Analysis of Clinical Manifestations and Different Diagnostic Methods of NeurocysticercosisSuspected Tamilian Population Residing in and Around Puducherry

Introduction: Human beings are accidental hosts for Cysticercus cellulosae showing varied clinical manifestations based on the site harbored by the parasitic cyst because of which disease profile remains unexplored at la...

Role of Ambulatory Blood Pressure Monitoring in Chronic Hypertensive Patients on Antihypertensive Therapy-A Cross-sectional Study

ABSTRACT Introduction: Regular monitoring of blood pressure in chronic hypertensive patients on antihypertensive therapy is essential to assess cardiovascular events and to prevant target organ damage. Aim: The present...

Triage of Children with Severe Acute Malnutrition and its Outcome: Single Centre Cross-sectional Study

Abstract Introduction: World Health Organisation (WHO), Emergency Triage, Assessment and Treatment (ETAT) guidelines categorised three most common presentations of sick children at admission; these are Respiratory distre...

Initial Experience with Grafalon as Induction Agent in Kidney Transplantation

ABSTRACT Introduction: Renal transplantation is ideal modality of Renal Replacement Therapy (RRT) as it is cost effective and associated with quality of life. Induction immunosuppression is an immunosuppressive therapy g...

Download PDF file
  • EP ID EP524005
  • DOI 10.7860/JCDR/2018/31225.11592
  • Views 71
  • Downloads 0

How To Cite

Anika Amritanand, Arathi Ro ddam Simha, Ringsar Baglari, Jenifer Roselin, John Michael, Gowri Mahasampath (2018). Agreement of Axial Length Measurements by Three Different Biometric Devices. Journal of Clinical and Diagnostic Research, 12(6), 1-4. https://europub.co.uk/articles/-A-524005