An Evaluation of the Effectiveness of the Structural Radiation Shielding Barriers of a Radiation Therapy Facility: Cancer Institute of Guyana

Journal Title: Journal of Medical Diagnostic Methods - Year 2018, Vol 7, Issue 1

Abstract

This project thoroughly investigates the integrity of the structural radiation shielding barriers of the radiation therapy facility of Cancer Institute of Guyana which houses a Linear Accelerator (LINAC) of 6 MV. The LINAC comprises a retractable beam stopper which renders all barriers of the bunker as secondary. However, the barriers that receive the primary incidence of the beam at the gantry angles 900 and 2700, Walls A and B, have been annotated as primary. In addition, for the other barriers, a phantom was used in making measurements for scatter. The bunker at Cancer Institute has been in existence since 2006. Therefore, the primary aim of this study is to validate the effectiveness of the structural radiation barriers of the radiation therapy facility through measurements and computations of the Instantaneous, Time Averaged and Weekly Dose Rates of the barriers and calculations of the attenuation coefficients using the Tenth value layer of the existing shielding material, concrete. The results gathered serve as a basis for comparative analysis to the Institute’s Commissioning Report and the International Standards. The Instantaneous measurements were recorded using the Fluke Biomedical Ionization Chamber. The results of the project annotate that the radiation shielding barriers at Cancer Institute of Guyana are very effective. The instantaneous dose rates recorded by the ionization chamber of the existing barriers were significantly lower than those calculated. The time averaged and weekly dose rates were very low as well. Scattered radiation measured conformed with the international standard of being less than 0.1% of the primary beam. However, calculations for barrier thicknesses required were higher than the existing barrier thicknesses. This was as a result of the field size parameter of the LINAC being utilized at its maximum. Nevertheless, these verifications of the effectiveness of the radiation shielding barriers were in conformity to the stipulated standards and well below calculated outcomes.

Authors and Affiliations

Parmeshwarie Seodat, Petal P Surujpaul, Deivis Errada

Keywords

Related Articles

Comparing a Polymerase Chain Reaction (PCR) Based Analysis for Gender Determination in Early Stage of Pregnancy versus Sonography

Objective: The aim of this study was to evaluate two genes as candidates for fetal sex determination by PCR from blood samples during the first trimester of pregnancy and then contrast the results with the sonography in...

Diagnostic Accuracy of 3T MR in Staging of Uterine Cervical Carcinoma

Objective: We aimed to evaluate the diagnostic accuracy of 3T MRI sequences for staging of uterine cervical carcinoma (CC).Methods: We examined 22 cases of CC by a 3T MRI with T2W single shot turbo spin echo, diffusion w...

An Evaluation of the Effectiveness of the Structural Radiation Shielding Barriers of a Radiation Therapy Facility: Cancer Institute of Guyana

This project thoroughly investigates the integrity of the structural radiation shielding barriers of the radiation therapy facility of Cancer Institute of Guyana which houses a Linear Accelerator (LINAC) of 6 MV. The LIN...

Leveraging Information Technology to Hardwire Diagnostic Stewardship

Objective: This article reports the impact of an electronic educational tool embedded in the electronic medical record on diagnostic stewardship of procalcitonin (PCT)-A biomarker used to differentiate bacterial from vir...

Value of Alberta Stroke Program early Computed Tomography Score (ASPECTS) in the Evaluation of Middle Cerebral Artery Acute Ischemic Stroke using CT and DW-MRI

Background: Alberta Stroke Program Early Computed Tomography Score (ASPECTS) is a quantitative score to evaluate early ischemic changes in the middle cerebral arterial territory on Computed Tomography (CT) in addition to...

Download PDF file
  • EP ID EP580845
  • DOI 10.4172/2168-9784.1000265
  • Views 97
  • Downloads 0

How To Cite

Parmeshwarie Seodat, Petal P Surujpaul, Deivis Errada (2018). An Evaluation of the Effectiveness of the Structural Radiation Shielding Barriers of a Radiation Therapy Facility: Cancer Institute of Guyana. Journal of Medical Diagnostic Methods, 7(1), 1-6. https://europub.co.uk/articles/-A-580845