Application of improved comparability verification method in performance verification of instruments and equipment
Journal Title: Chinese Journal of Blood Transfusion - Year 2022, Vol 35, Issue 10
Abstract
Objective To explore the viability of improved comparability verification method for the performance verification of automatic biochemical analyzer in blood centers. Methods According to the relevant requirements of User Verification of Performance for Precision and Trueness; Approved Guideline-Second Edition (EP15-A2) and Guideline for Comparability Verification of Quantitative Results with One Health Care System(WS/T 407-2012), the performance of the newly purchased Beckman Coulter AU480 automatic biochemical analyzer was confirmed by means of comparability verification, and compared with the current Olympus AU400 automatic biochemical analyzer. Results The detection results given by Beckman AU480 biochemical analyzer on 5 samples for ALT detection performance test were significantly correlated with the target value (R2=0.998), and the trueness verification results met the requirements.The CV of medium and high ALT levels was 2.36% and 1.20%, respectively, within the acceptable standard range(<5%), and the precision verification results meet the requirement.The comparison deviation of two devices for medium and high ALT levels was 0.47% and 1.62%, respectively, within the acceptable standard range (<5%). The detection results of ALT from 40 blood donors given by two devices were significantly correlated (R2=0.956), and the precision verification results met the requirements.In the validation of linear evaluation, there was a good linear relationship between the expected and measured results (r=0.997)among samples with 6 levels of concentrations. Conclusion The comparability verification method is suitable for the performance verification of automatic biochemical analyzer in blood centers.
Authors and Affiliations
Jiang QIAN, Zhongmei LIU, Xiaomin ZHOU
Effects of APOBEC3G on the replications of severe fever with thrombocytopenia syndrome virus
Objective To study the effect of apolipoprotein B mRNA editing enzyme catalytic subunit 3G (APOBEC3G)on virus replication and IL-6 production during severe fever with thrombocytopenia syndrome virus (SFTSV) infection. Me...
Effect of applied muscle tension on improving vasovagal reactions among blood donors: a meta-analysis
Objective To systematically evaluate the effect of applied muscle tension (AMT) exercises on reducing vasovagal reactions(VVRs) among blood donors by meta-analysis. Methods Nine related databases including MEDLINE, Web o...
Analysis of chylemia and its impact on blood product quality
Objective To analyze the causes of chylemia, explore the effects of varying degrees of chylemia on subsequent blood products and the countermeasures. Methods The information of blood donors in the Central Blood Station o...
Clinical application of thrombelastogram and reference interval for our laboratory
Objective To respectively analyze the application of thrombogram (TEG) in our hospital and explore the reference interval for our laboratory. Methods Healthy adults who underwent physical examination in our hospital from...
Factors influencing the returning of plateletpheresis donors
Objective To explore and analyze on the factors that influence the returning of plateletpheresis donors. Methods A total of 1 302 plateletpheresis donors, selected from 2014 to 2020 in the blood center, were followed up...