Characterization Of Structures Of Equivalent Tissue With a Pixel Detector

Abstract

Research using hybrid pixel detectors in medical physics is on the rise. Timepix detectors have arrays of 256 × 256 pixels with a resolution of 55 µm. Here, and by using Timepix counts instead of Hounsfield units, we present a calibration curve of a Timepix detector analog to those used for CT calibration. Experimentation consisted of the characterization of electron density in 10 different kinds of tissue equivalent samples from a CIRS 062M phantom (lung, 3 kinds of bones, fat, breast, muscle, water and air). Radiation of the detector was performed using an orthodontic X-ray machine at 70 KeV and .06 second of tube current with a purpose-built aluminum collimator. Data acquisition was performed at 1 frame per second and taking 3 frames per phantom. We were able to find a curve whose behavior was similar to others already published. This will lead to the verification of the usage of Timepix for identification of different tissues in an organ.

Authors and Affiliations

M. C. Grados Luyando, B De Celis-Alonso, E. Moreno-Barbosa, M. I. Martinez Hernandez, J. M. Hernandez Lopez, G. Tejeda Munoz

Keywords

Related Articles

Elemental Analysis of Nanomaterial Using Photon-Atom Interaction Based EDXRF Technique

Presence of trace amount of foreign impurities (both metallic and non-metallic) in standard salts used for sample preparation and during the synthesis process can alter the physical and chemical behavior of the pure and...

The Status of Natural Radioactivity and Heavy Metals Pollution on Marine Sediments Red Sea Coast, At Safaga, Egypt

Natural 226Ra, 232Th and 40K radio nuclides concentration in beach Sediments along Safaga coast of Red sea, Egypt has been carried out using a NaI (Tl) gamma ray spectrometric technique. The total average concentrations...

Nucleosynthesis in Neutron Stars Crust

We address the problem of the origin of heavier and super heavy elements in nature. Believing rapid neutron capture process (r-process) is responsible for synthesis of heavy and super heavy elements in the remnants of su...

ZnS (Mn) Nanoparticles as Luminescent Centers for Siloxane Based Scintillators

Synthesis of oleic acid stabilized ZnS nanocrystals activated with Mn is pursued. A hydrothermal method where high pressure and temperature are applied to control the nanocrystals growth is adopted. Capping the nanoparti...

Goodness of Generalized Seniority in Even-even Sn Isotopes

Seniority has proved to be a unique and simple probe to address some of the complex issues underlying nuclear structure of nuclei close to magic numbers. An extension from the concept of seniority in single-j shell to ge...

Download PDF file
  • EP ID EP595705
  • DOI 10.15415/jnp.2017.51008
  • Views 203
  • Downloads 0

How To Cite

M. C. Grados Luyando, B De Celis-Alonso, E. Moreno-Barbosa, M. I. Martinez Hernandez, J. M. Hernandez Lopez, G. Tejeda Munoz (2017). Characterization Of Structures Of Equivalent Tissue With a Pixel Detector. Journal of Nuclear Physics, Material Sciences, Radiation and Applications, 5(1), 79-89. https://europub.co.uk/articles/-A-595705