Characterization Of Structures Of Equivalent Tissue With a Pixel Detector

Abstract

Research using hybrid pixel detectors in medical physics is on the rise. Timepix detectors have arrays of 256 × 256 pixels with a resolution of 55 µm. Here, and by using Timepix counts instead of Hounsfield units, we present a calibration curve of a Timepix detector analog to those used for CT calibration. Experimentation consisted of the characterization of electron density in 10 different kinds of tissue equivalent samples from a CIRS 062M phantom (lung, 3 kinds of bones, fat, breast, muscle, water and air). Radiation of the detector was performed using an orthodontic X-ray machine at 70 KeV and .06 second of tube current with a purpose-built aluminum collimator. Data acquisition was performed at 1 frame per second and taking 3 frames per phantom. We were able to find a curve whose behavior was similar to others already published. This will lead to the verification of the usage of Timepix for identification of different tissues in an organ.

Authors and Affiliations

M. C. Grados Luyando, B De Celis-Alonso, E. Moreno-Barbosa, M. I. Martinez Hernandez, J. M. Hernandez Lopez, G. Tejeda Munoz

Keywords

Related Articles

Partial as Well as Total Photon Interaction Effective Atomic Numbers for Some Concretes

Photon interaction effective atomic number (Zeff) for partial as well as total photon interaction processes has been computed using logarithmic interpolation method for seven different concretes viz. (i) Ordinary, (ii) H...

Stability of Aspartic Acid at 77°K under Gamma Radiation in a Comet Cores Simulation: Implications for Chemical Evolution Studies

The synthesis of organic matter in a simulated primitive environment (terrestrial or extraterrestrial) has been widely studied. The stability of organic matter of biological significance, exposed to energy fields in prim...

Fission in Rapidly Rotating Nuclei

We study the effect of rotation in fission of the atomic nucleus256Fm using an independent-particle shell model with the mean field represented by a deformed Woods-Saxon potential and the shapes defined through the Cass...

Alignment Studies for Tungsten Near L3 Sub-Shell Threshold Via Theoretical, Experimental and Empirical Methods

Alignment studies are made for tungsten near L3 sub-shell threshold using theoretical, experimental and empirical approaches. Experimentally to measure alignment parameter, the angular distribution of L x-rays of tungste...

Studies on Neutron, Photon (Bremsstrahlung) and Proton Induced Fission of Actinides and Pre-Actinides

We present the yields of various fission products determined in the reactor neutron, 3.7-18.1 MeV quasi-mono energetic neutron, 8-80 MeV bremsstrahlung and 20-45 MeV proton induced fission of 232Th and 238u using radioch...

Download PDF file
  • EP ID EP595705
  • DOI 10.15415/jnp.2017.51008
  • Views 222
  • Downloads 0

How To Cite

M. C. Grados Luyando, B De Celis-Alonso, E. Moreno-Barbosa, M. I. Martinez Hernandez, J. M. Hernandez Lopez, G. Tejeda Munoz (2017). Characterization Of Structures Of Equivalent Tissue With a Pixel Detector. Journal of Nuclear Physics, Material Sciences, Radiation and Applications, 5(1), 79-89. https://europub.co.uk/articles/-A-595705