COIL DESIGN WITH LITZE WIRE FOR MAGNETIC PARTICLE SPECTROMETRY

Abstract

The design of an excitation coil for magnetic particles spectrometer (MPS) was described. It was assumed that the spectrometer should measure the spectra of particles of diameter in the range 10-100 nm. To measure the amplitude and phase angle spectra of magnetic nanoparticles it is required to generate sinusoidal alternating spatially homogeneous magnetic field of magnitude of 20 mT. The work volume of the designed spectrometer was 202020 mm allowing measurement of small samples. The estimation of magnetic properties of magnetic nanoparticles is crucial in Magnetic Particles Imaging. In this paper we described the excitation coil design which minimizes power losses on the coil related to heat emission. The resonance circuit operating on 20 kHz frequency was applied. Optimal litze wire configuration was proposed to negate skin and proximity effects. The numeric simulations for the optimal and commercially available suboptimal litze wire configurations were performed. The comparison of the results was shown and discussed.

Authors and Affiliations

Przemysław Wróblewski, Waldemar Smolik

Keywords

Related Articles

PORÓWNANIE NUMERYCZNYCH MODELI PORTÓW LINII PASKOWEJ

Celem artykułu jest przedstawienie różnych sposobów modelowania portów struktur takich jak linie paskowe albo anteny. Przedstawionoteż praktyczną realizację opisywanych modeli, wykonaną przy użyciu pakietu COMSOL. Przean...

INVERSE PROBLEM SOLUTION USING NEURAL NETWORK

This paper was presented a practical method of solving inverse problems in tomography using neural networks. The methods to detect changes in the size and position of an object inside another object tomography. Computed...

ROZMYTY SYSTEM ROZPOZNAJĄCY I STERUJĄCY PRZEPŁYWAMI DWUFAZOWYMI MIESZANIN GAZ-CIECZ

W artykule opisano system rozmyty do rozpoznawania przepływów dwufazowych mieszanin gaz-ciecz w oparciu o trójwymiarowe dane tomograficzne oraz informacje kontekstowe. System dedykowany jest również do sterowania pracą i...

IDENTYFIKACJA BAKTERII W WIDMIE BLISKIEJ PODCZERWIENI

Identyfikacja mikroorganizmów, głównie identyfikacja bakterii i wykrywanie patogenów jest niezwykle istotnym zagadnieniem w wielu dziedzinach mikrobiologii takich jak: diagnozowanie infekcji czy ochrona żywności. W tym a...

BEZPIECZEŃSTWO STOSOWANIA LAMP LED

Przedstawiono doniesienia literaturowe oraz własne wyniki badań promieniowania optycznego lamp LED i oceniono stwarzane przeznie zagrożenie fotobiologiczne w oparciu o normę PN-EN 62471 „Bezpieczeństwo lamp i systemów la...

Download PDF file
  • EP ID EP221809
  • DOI 10.5604/01.3001.0010.4605
  • Views 71
  • Downloads 0

How To Cite

Przemysław Wróblewski, Waldemar Smolik (2017). COIL DESIGN WITH LITZE WIRE FOR MAGNETIC PARTICLE SPECTROMETRY. Informatyka Automatyka Pomiary w Gospodarce i Ochronie Środowiska, 7(1), 150-153. https://europub.co.uk/articles/-A-221809