COMPARISON BETWEEN SIMPLE POLYNOMIALS AND AN ITS-90 BASED REFERENCE FUNCTIONS OF EMF VS. TEMPERATURE RELATION FOR HIGH PURITY NOBLE METAL THERMOCOUPLES

Journal Title: Metrology and Measurement Systems - Year 2005, Vol 12, Issue 2

Abstract

Emf versus temperature reference polynomial functions covering the temperature range from 100° C up to 960° C have been developed for three types of noble metal thermocouples, which are Pt-10%Rh/Pt, Au/Pt and Pt/Pd. Our functions are compared to the reference functions available in the literature. The thermocouples were fabricated and annealed for stabilization at 962° C for 300 hours before calibrations. After stabilization they were calibrated by comparison with Standard Platinum Resistance Thermometer (SPRT) in the temperature range from 100° C to 450° C and by comparison with a High temperature Standard Platinum Resistance Thermometer (HSPRT) in the temperature range from 450° C up to 950° C. The error of fit was evaluated and discussed.

Authors and Affiliations

FATEN MEGAHED

Keywords

Related Articles

DIGITAL-TO-ANALOG CONVERSION USING DELTA-SIGMA MODULATORS IMPLEMENTED IN VIRTEX-4 FPGA DEVICES

This paper presents some selected architectures of delta-sigma modulators which are suitable for implementation of digital-to-analog converters (DACs) in modern FPGA devices. Simulation and implementation of simple archi...

IN MEMORY OF THE PROFESSOR DR HAB. BRUNON SZADKOWSKI

On February 2nd, 2007 Professor Brunon Szadkowski, an eminent scientist and meritorious research worker of the Faculty of Electrical Engineering of the Silesian University of Technology of Gliwice, departed this world.

PARTICULARITIES OF THE CYCLIC A/D CONVERTERS ENOB DEFINITION AND MEASUREMENT

The paper is devoted to the methods of accurate assessment and analysis of the expected and actual effective number of bits (ENOB) of cyclic analog-to-digital converters. The equivalence of definition of ENOB in the IEEE...

INTRODUCTION TO THE T AND CHI-SQUARE DISTRIBUTION FOR A MORE ACCURATE EVALUATION OF THE MEASURE OF THE WORD ERROR RATE IN ANALOGTO-DIGITAL CONVERTERS<br />

The word error rate (WER) in an Analog to Digital Converter (ADC) is the probability of receiving an erroneous code for an input, after correction is made for gain, offset, and nonlinearity errors, and a specified allowa...

DIGITAL MEASUREMENT OF PHASE DIFFERENCE - A COMPARATIVE STUDY OF DSP ALGORITHMS

The paper compares nine methods of measurement of the phase difference of digitized sinusoidal signals mainly according to their sensitivity to sampling non-integer number of signal periods. The investigated methods can...

Download PDF file
  • EP ID EP118661
  • DOI -
  • Views 86
  • Downloads 0

How To Cite

FATEN MEGAHED (2005). COMPARISON BETWEEN SIMPLE POLYNOMIALS AND AN ITS-90 BASED REFERENCE FUNCTIONS OF EMF VS. TEMPERATURE RELATION FOR HIGH PURITY NOBLE METAL THERMOCOUPLES. Metrology and Measurement Systems, 12(2), 165-174. https://europub.co.uk/articles/-A-118661