Comparison of corneal curvature and astigmatism measurements with four kinds of instruments before cataract surgery

Journal Title: Guoji Yanke Zazhi - Year 2018, Vol 18, Issue 5

Abstract

AIM: To compare corneal curvature and astigmatism values identified with OPD-Scan Ⅲ, IOL Master, Pentacam and KR-8900 in cataract patients and to evaluate the consistencies and differences. <p>METHODS: In this prospective clinical trial, 45 cataract patients(86 eyes), received treatment at the Department of Ophthalmology in our hospital from April 2017 to June 2017, underwent corneal parameters measurements with four instruments: OPD-Scan Ⅲ, Pentacam, IOL Master, and KR-8900. Measurements included corneal curvature(K<sub>1</sub>, K<sub>2</sub>, K<sub>m</sub>)and astigmatism parameters. The results were assessed using the paired samples <i>t</i> tests, Pearson coefficient, and Bland-Altman plots. <p>RESULTS: The difference of K<sub>1</sub> value measured by OPD-Scan Ⅲ and IOL Master was significant(<i>P</i><0.001). Significant differences were also found in K<sub>2</sub> and K<sub>m</sub> values between OPD-Scan Ⅲ and Pentacam or IOL Master(<i>P</i><0.05), while others had noncomparable results(<i>P</i>>0.05). All biometry measurements revealed good liner correlation(<i>r</i>=0.408-0.980, <i>P</i><0.001). The Bland-Altman showed that the corneal curvature and astigmatism vectors examined by OPD-Scan Ⅲ and all other devices had non-comparable results, except for the comparison of K<sub>1</sub>, K<sub>2</sub> and K<sub>m</sub> measured by OPD-Scan Ⅲ and KR-8900. <p>CONCLUSION: OPD-Scan Ⅲ and KR-8900 can be interchangeable for the measurements of corneal curvature, while the others are not directly interchangeable. The astigmatic results obtained from OPD-Scan Ⅲ are not completely interchangeable with KR-8900, IOL Master or Pentacam.

Authors and Affiliations

Xue-Xi Li

Keywords

Related Articles

Effect of calcium dobesilate on fundus microcirculation in patients with diabetic retinopathy and cataract

AIM: To analyze the effect of calcium dobesilate combined with hypoglycemics for patients with diabetic retinopathy(DR)and cataract, and the influence on microcirculation of eye fundus, hypoxia inducible factor-1α(HIF-1α...

Analysis of the refraction error and the influencing factors after phacoemulsification in acute primary angle-closure glaucoma with cataract

AIM: To compare the refraction error(RE)3mo after phacoemulsification combined with intraocular lens implantation(PHACO+IOL)between patients with acute primary angle-closure glaucoma(APACG)with cataract(APACG group)and p...

Effect of built-in type OASIS iris expander in phacoemulsification surgery for cataract with small pupil

AIM: To observe the effect of built-in type OASIS iris expander in phacoemulsification surgery for cataract with small pupil. <p>METHODS: Totally 44 cataract patients with small pupil were treated in our hospital from Ja...

Correlation between symptoms and sign of drug users with dry eye

AIM: To investigate the correlation between symptoms and sign of drug users with dry eye. <p>METHODS: Eighty-seven patients(87 eyes)with dry eye syndrome who were abstained from a compulsory rehabilitation center in Yang...

Anti-VEGF medicine with PRP for neovascular glaucoma

AIM:To investigate the effect of anti-vascular endothelial growth factor(VEGF)combined with pan retinal photocoagulation(PRP)on neovascular glaucoma(NVG)and its effect on VEGF and platelet-derived growth factor-C(PDGF-C)...

Download PDF file
  • EP ID EP656687
  • DOI 10.3980/j.issn.1672-5123.2018.5.34
  • Views 95
  • Downloads 0

How To Cite

Xue-Xi Li (2018). Comparison of corneal curvature and astigmatism measurements with four kinds of instruments before cataract surgery. Guoji Yanke Zazhi, 18(5), 904-908. https://europub.co.uk/articles/-A-656687