Consistency evaluation of eyeball biological measurements using StarEyes 900 and IOLMaster 500
Journal Title: Eye Science (Yanke Xuebao) - Year 2022, Vol 37, Issue 2
Abstract
[Objective:] To evaluate the difference, correlation and agreement of eye parameters measured by StarEyes 900 visual function analyzer (Wan Ling Bang Qiao, China) and IOLMaster 500 (Carl Zeiss, Germany) swept-source optical coherence tomography biometer. [Methods:] A prospective study was designed involving 62 healthy subjects (124 eyes) undergoing ophthalmic examinations in Zhongshan Ophthalmic Center from June 2021 to July 2021. Data from their both eyes were selected for analysis in all patients. Axial length (AL), keratometry for the steepest meridian (Ks), keratometry for the flattest meridian (Kf), mean keratometry (Km) and corneal diameter (WTW) were measured by the StarEyes 900 visual function analyzer and IOLMaster 500 swept-source optical coherence tomography biometer. A paired t-test was used to analyze the differences in measurement results. The Pearson correlation coefficient was used to analyze the correlation. Bland-Airman method was used to assess the agreement of the instruments. [Results:] The AL, Kf, Ks, Km and WTW obtained by StarEyes 900 and IOLMaster 500 were (24.18±1.08) mm and (24.16±1.08) mm, (42.84±1.65) D and (43.04±1.57) D, (44.34±1.90) D and (44.17±1.80) D, (43.59±1.73) D and (43.61±1.64) D, and (11.64±0.29) mm and (11.64±0.30) mm, respectively. The Km and WTW of the two devices showed no significant difference (P>0.05), while the AL, Ks and Kf showed significant differences (all P<0.01). The AL and Ks obtained by StarEyes 900 were higher than by IOLMaster 500, while the Kf, Km and WTW were lower. The measurements of five aforementioned biometric parameters by both devices showed good correlation by Pearson correlation coefficient and good agreement by Bland-Airman. [Conclusion:] The Km and WTW measured by the two devices showed no significant difference, and provided references to one another. The difference in AL, Kf and Ks between the two devices showed significant differences. All of the measurements showed good correlation by Pearson correlation coefficient and good agreement by Bland-Airman.
Authors and Affiliations
Yuxuan WU, Zhuoling LIN, Xiaohang WU, Pisong YAN, Haotian LIN
Exploration and practice in optometry education mode under the background of visual health management
The World Health Organization and Chinese “Healthy China” strategy have elevated visual health management to an unprecedented height. Ensuring people's visual health multi-directionally in whole period and enhancing peop...
Analysis of influencing factors and General Well-Being Schedule evaluation in early-postoperative cataract patients during the COVID-19 epidemic
[Objective:] To evaluate the general well-being and its associated factors for the early-postoperative patients with different types of cataract during the epidemic of corona virus disease 2019 (COVID-19), and to provide...
Health belief and influencing factors of ophthalmic follow-up in patients with diabetes in medical alliance
[Objective:] To measure the adherence to ophthalmology follow-up and the related health belief of patients with diabetes in medical alliance, and analyze the influencing factors. [Methods:] Totally 334 community-dwelling...
Urban particulate matter related ocular surface diseases and their potential pathogenic mechanism
More and more evidence indicate that air pollution can seriously damage human health and become a major cause of the global disease burden. The main component of air pollution, particulate matter, can penetrate the lungs...
Differences in pediatric ophthalmology and strabismus between China and the U.S.
This article briefly introduced the discipline construction and clinical diagnosis and treatment routines of pediatric ophthalmology and strabismus at the University of California, San Francisco, describing the major dif...