Consistency of two eye anterior segment analysis systems measuring wave-front aberration before and after refractive surgery
Journal Title: Guoji Yanke Zazhi - Year 2018, Vol 18, Issue 11
Abstract
AIM: To study the consistency of pre-operative and postoperative wave-front aberration between SCHWIND and SIRIUS eye anterior segment analysis system. <p>METHODS: Prospective control study. A total of 360 eyes of 180 patients with refractive errors in West China Hospital of Sichuan University were selected. The data, including the degree and the direction of the flat axis and the steep axis, Kappa angle and a variety of high order wave-front aberrations, were measured by SCHWIND and SIRIUS anterior segment analyzers. The average value of each parameter was taken 5 times, and paired sample <i>t</i> test was used and the data was analyzed by SPSS22.0 software. <p>RESULTS: Flat axis degree and direction, steep axis degree and direction, coma, and trefoil were measured preoperatively and 1mo postoperatively by SCHWIND and by SIRIUS eye anterior segment analysis systems, and there were no statistically significant differences between the two machines(<i>P</i>>0.05). Kappa angle was measured 1mo postoperatively and there was no statistically significant differences(<i>P</i>>0.05). There was statistically significant difference(<i>P</i><0.01)preoperatively between two machines of high-order aberrations and spherical aberration(SpAb); values measured by SCHWIND were higher than those measured by SIRIUS. There was no statistically significant difference(<i>P</i>>0.05)between two machines of high-order aberrations or SpAb in 1mo postoperatively. <p>CONCLUSION: SCHWIND and SIRIUS anterior segment eye analysis system have a good consistency of wave-front aberration measurement before and after refractive surgery.
Authors and Affiliations
Xiao-Lan Zhang
Effect and safety of EX-PRESS shunt implantation combined with biological amniotic membrane implantation for primary open angle glaucoma
AIM: To analyze the curative effect and safety in early stage primary open angle glaucoma(POAG)between simple Ex-PRESS shunt implantation and its combination with biological amniotic membrane implantation. <p>METHODS: Th...
Comparison of wide-angle digital retinal imaging system and binocular indirect ophthalmoscope in retinopathy of prematurity screening
AIM: To compare the application effect on patients with retinopathy of prematurity(ROP)screened by wide-angle digital retinal imaging system(RetCam)and binocular indirect ophthalmoscope(BIO). <p>METHODS: Totally 1 624 ca...
Effect of intravitreal injection of Conbercept before PPV on complications and visual recovery in patients with PDR
AIM: To analyze the effect of intravitreal injection of Conbercept before pars plana vitrectomy(PPV)on complications and visual recovery in patients with proliferative diabetic retinopathy(PDR). <p>METHODS: Totally 94 pa...
Effect of long-term wearing rigid gas permeable contact lenses on the stability of tear film
AIM: To investigate the effect of long-term wearing rigid gas permeable(RGP)contact lenses on the stability of tear film, and to observe its safety. <p>METHODS: Totally 60 myopia patients(120 eyes)wearing RGP in our hosp...
Clinical efficacy of disinhibition and visual function training in children patients with anisometropic amblyopia
AIM:To evaluate the clinical efficacy of treatment <i>via</i> disinhibition and visual function training in children patients with anisometropic amblyopia. <p>METHODS: Eighty-three children patients with anisometropic am...