CORRELATION OF INTRAOCULAR PRESSURE MEASUREMENTS WITH NON CONTACT TONOMETER AND GOLDMANN APPLANATION TONOMETRY

Journal Title: Journal of Evidence Based Medicine and Healthcare - Year 2017, Vol 4, Issue 56

Abstract

BACKGROUND A complete ophthalmologic examination includes intraocular pressure (IOP) measurement, which is a routine procedure and is important for diagnosis and monitoring of glaucoma. IOP measurement is most commonly done using Goldmann Applanation tonometer and Non-Contact tonometer. MATERIALS AND METHODS In this study IOP measurements of 500 eyes (glaucomatous and non-glaucomatous) were performed using GAT and NCT on patients visiting the outpatient clinic of Department of Ophthalmology at Christian Medical College and Hospital, Ludhiana. This was a cross sectional and observational study. Comparison of IOP values was done in different IOP ranges. CCT was measured and analysis of its correlation with GAT and NCT was done. RESULTS Both methods of tonometry correlated significantly in patients with IOP <24 mm Hg. The mean IOP measured by NCT, was 16.06 ± 5.85 mm Hg and the mean IOP measurement by GAT was 16.61 ± 6.94 mm Hg. Intraocular pressure readings with GAT and NCT positively correlated with CCT. CONCLUSION NCT may be useful for screening in clinical settings but borderline high IOP readings should be confirmed with GAT. Our findings, also suggest that CCT is an essential variable to consider in interpreting IOP readings.

Authors and Affiliations

Leya Sara George, Arun Bhatti, Nitin Batra

Keywords

Related Articles

A STUDY ON THE ROLE OF MICRO NUCLEI IN ASSESSING THE PROGRESSION OF PRECANCEROUS LESIONS OF CERVIX AND THE DIAGNOSIS OF CARCINOMA OF CERVIX

BACKGROUND: Invasive squamous cell carcinoma of cervix remains the most common malignant neoplasm of the female genital tract in many countries. The Papanicolaou stained cervical smear is an excellent and effective metho...

PHARMACOLOGY OF NEWER ANTIMALARIAL DRUGS: REVIEW ARTICLE

Malaria is currently is a major health problem, which has been attributed to wide spread resistance of the anopheles mosquito to the economical insecticides and increasing prevalence of drug resistance to plasmodium falc...

ROLE OF THE HEALTH-CARE PROFESSIONALS IN COMBATING TOBACCO EPIDEMICS

Tobacco epidemics have become a matter of serious health hazard all over the world. Smoking tobacco considered as a means for relaxation and recreation is now indicated as a menacing health hazard. Tobacco in all forms a...

COMPARATIVE STUDY OF COMPLICATIONS IN ENDOSCOPIC SEPTOPLASTY VERSUS CONVENTIONAL SEPTOPLASTY

BACKGROUND The objectives of the study are- 1) To study the gender distribution of Deviated Nasal Septum (DNS). 2) To study the incidence of DNS in different age groups. 3) To compare the complications of septoplasty in...

A SURVEY ON THE INCIDENCE OF PSEUDO-THROMBOCYTOPENIA IN THE IN-PATIENTS

Fever with thrombocytopenia is frequently encountered in the clinical settings. However some cases of thrombocytopenia are associated with clumping of platelets resulting in pseudothrombocytopenia. This results in dispar...

Download PDF file
  • EP ID EP227491
  • DOI 10.18410/jebmh/2017/683
  • Views 81
  • Downloads 0

How To Cite

Leya Sara George, Arun Bhatti, Nitin Batra (2017). CORRELATION OF INTRAOCULAR PRESSURE MEASUREMENTS WITH NON CONTACT TONOMETER AND GOLDMANN APPLANATION TONOMETRY. Journal of Evidence Based Medicine and Healthcare, 4(56), 3430-3433. https://europub.co.uk/articles/-A-227491