Diagnostic Value of Digital Radiography in Gap Detection of Implant-Abutment Connection with Zirconia Abutments in Different Vertical X-ray Projection Angles
Journal Title: Journal of Clinical and Diagnostic Research - Year 2018, Vol 12, Issue 5
Abstract
ABSTRACT Introduction: Accurate seating of abutments on implants is an important factor for long term success of implant supported restorations. Among different methods of evaluating abutment seating, radiography is most commonly used, while some clinical factors such as angulation of radiography film and radiography tube can adversely affect its diagnostic value. Aim: To evaluate diagnostic value of digital periapical radiographs in terms of sensitivity, specificity and accuracy of gap detection at implant-abutment connection with different angles of X-ray projection when using a zirconia abutment. Materials and Methods: In this in vitro study radiographs were made of three different implant connections (internal tri-channel, external hex and internal hex with platform switching) to which zirconia abutments were fastened, once with and once without using a spacer between implant and abutment. Radiographs were taken at different projection angles of -20, -15, -10, -5, 0, +5, +10, +15, and +20. Then 10 prosthodontists were asked to assess all the radiographs, two times for each one with an interval of 10 days. Sensitivity, specificity and accuracy indices were calculated for each system. Results: Findings showed that sensitivity of gap detection in all study groups was amongst acceptable range (85-100%). In general external hex connection system showed better results (95-100%) while the platform switching internal hex system had the weakest results (85-100%). Conclusion: Angle of X-ray projection is better to be between -10 to +10 when evaluating complete seat of zirconia abutments on implants especially if a platform switching system is used.
Authors and Affiliations
Marzieh Alikhasi, Alireza Salehi, Maryam Memarian, Mahtab Tabesh
Women’s Views on Routine Antenatal Ultrasound Scan in a Low Resource Nigerian Setting
ABSTRACT Introduction: Obstetric ultrasound is a popular and attractive management tool in Obstetrics. In recent times, it has gained wide acceptance amongst expectant mothers in many parts of the developing world. Many...
Correlation of Trace Elements in Patients of Chronic Liver Disease with Respect to Child- TurcottePugh Scoring System
Introduction: Zinc, copper, manganese and magnesium are essential trace elements whose role in chronic liver disease and its complications is not clear. Aim: To study the concentration of these elements in patients with...
Depression, its Correlates and Effects in Ever Married Urban Women Residing in Kumaon Region of Uttarakhand
ABSTRACT Introduction: In the global context, independent studies on depression show that women are twice as likely as men to suffer from clinical depression. It is also described as an important cause of premature morta...
Anaesthetic Considerations for Balloon Mitral Valvuloplasty in Pregnant Patient with Severe Mitral Stenosis: A Case Report and Review of Literature
Even though, the incidence of Mitral Stenosis (MS) has reduced grossly, it still remains a health problem in developing countries and is the most common rheumatic valvular lesion encountered in pregnant patients. The alr...
Morphometric Study of Foramina Transversaria and the Incidence of Accessory Foramina in Cervical Spine of Indian Population
ABSTRACT Introduction: The Foramina Transversaria (FT) are known to exhibit variations in their size, shape and they may be absent or duplicated/triplicated. The deformation of foramina may result in derangement of neuro...