Dose distributions verification for High dose rate brachytherapy plans by using ionization chambers 2D array

Journal Title: INTERNATIONAL JOURNAL OF COMPUTERS & TECHNOLOGY - Year 2016, Vol 15, Issue 3

Abstract

The purpose of this paper is to perform dosimetric verification (in-phantom) of dose distributions calculated with treatment planning system (TPS) by using  2D chamber array device in HDR brachytherapy. HDR brachytherapy treatment plans’dose distribution verification is performed using the two-dimensional (2D) ionization chamber array MatriXX Evolution developed by IBA Dosimetry (IBA Dosimetry, Germany) whose detector area is covered with the Nucletron Freiburg Flap Applicator Set (Nucletron BV,Veenendaal, the Netherlands) with catheters such that the detector plane was set to 0.86 cm from the catheter plane. Fixed slabs of RW3 (Perspex) were added below the 2D-ARRAY to provide full scattering conditions. The phantom was scanned on computed tomography (CT) for treatment planning with 2.5-mm slice thickness. Based on the CT data of the phantom, three different plans were calculated by the planning system (Oncentrabrachy version 4.3, Nucletron BV) and then are exported to the VERISOFT software for comparison with measured data. For comparison of dose distributions, both dose planes – measured & calculated – were normalized  to the global maximum dose of the reference matrix (measured data set) and compared using the Gamma index method. Gamma indexes were evaluated using a dose-difference criterion of 3% and a distance criterion of 3 mm (γ≤1).The total number of evaluated dose points for the vault test case  is 9755, 98.6 % of them (9623 point) passed the criteria of acceptability (3% delta dose and 3-mm distance criteria) and 1.4% of them (132 point) failed it. The total number of evaluated dose points for the full test case is 19964, 93.6 % of them (18683 point) passed  the criteria of acceptability and 6.4% of them (1281 point) failed it. And the total number of evaluated dose points for the cylinder test case is 19871, 96.9 % of them (19258 point) passed the criteria of acceptability and 3.1% of them (613 point) failed it. By thisThe use of the two-dimensional (2D) ionization chamber array MatriXX Evolution for brachytherapy applications has been successfully demonstrated. The array measurements in these planes have shown acceptable agreement with the TPS, generally within 3% delta dose and 3-mm distance agreement criteria within each plane. The comparisons made led to make a relation between the passing percentage values to the number of the evaluated dose points for each test case and it was found that as the number of these pixels increases the possibility of having more failing points increases.

Authors and Affiliations

Rasha Moustafa Abdelfattah, N. Ahmed Deiab, M. Hassan Elnaggar, M. Hany Khedr, R. Abdelmoneim Rizk

Keywords

Related Articles

Predilection of Reusability over Maintainability in Aspect-Oriented Systems

Maintenance is the important phase in software development lifecycle which initiates after the software has been deployed for use. Reusability is an important area of concern which depicts the extent to which a module c...

Modeling an Object Oriented for Maintenance Purposes

Software maintenance is the last phase of the software life cycle. The aim of the software maintenance is to maintain the software system in accordance with advancement in software and hardware technology. In this paper,...

Evaluation of economic advantage for produced electrical energy usage by solar cell in compression with steam power plant in home in Isfahan city

The aim of this article is evaluation of economic advantage from using of solar energy (photovoltaic system) in compression with steam power plant in order to supply the electricity of Isfahan’s household (2009 to 2011...

Using mobile communication as a tool for national growth and development

Mobile communication emergence has brought a whole palette of new possibilities in information sharing not only for individual development but also for national growth. New mobile information devices such as Personal Dig...

DEFINING QUALITY CRITERIA IN SITUATIONAL METHOD ENGINEERING

Method engineering is a field of engineering that has been created a need expressed by engineers. These engineers faced problems in using conventional methods such as SSADM, DSDM, OOAD...In this way, Method engineering w...

Download PDF file
  • EP ID EP650811
  • DOI 10.24297/ijct.v15i3.1685
  • Views 62
  • Downloads 0

How To Cite

Rasha Moustafa Abdelfattah, N. Ahmed Deiab, M. Hassan Elnaggar, M. Hany Khedr, R. Abdelmoneim Rizk (2016). Dose distributions verification for High dose rate brachytherapy plans by using ionization chambers 2D array. INTERNATIONAL JOURNAL OF COMPUTERS & TECHNOLOGY, 15(3), 6618-6625. https://europub.co.uk/articles/-A-650811