Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol

Journal Title: Journal of Biomedical Physics and Engineering - Year 2019, Vol 9, Issue 4

Abstract

Background: The commissioning accuracy of Volumetric Modulated Arc Therapy (VMAT) need to be evaluated. Objective: To test and evaluate commissioning accuracy of VMAT based on the TG 119 protocols at local institution. Material and Methods: The phantom, structure sets, VMAT and IMRT beam parameter setup, dose prescriptions and planning objectives were following TG 119 guidelines to create local treatment plans of VMAT and IMRT. The local planning results were compared with the results of TG 119. Point measurement at high and low dose regions were measured using three ionization chambers with different active volumes (CC01, CC13, FC65G). The composite dose was measured by a 2D detector array and analyzed for the percentage of points passing the gamma criteria of 3 % dose difference (DD) and 3 mm distance-to-agreement (DTA) and 2 % DD and 2 mm DTA. Results: The local treatment plans of VMAT and IMRT capable to meet the dose goals criteria set by TG 119 except for C-shape hard. Three ionization chambers with various active volumes for point measurement showed an increase in the confidence limit (CL), the larger the active volume was found proportional to increase the value of CL. The results obtained from ion chambers CC01 and CC13 could met the dose criteria set by TG 119, but results obtained from ion chamber FC65G fail the criteria. All gamma evaluation results show more than 95% data points pass the criteria of 3% DD and 3 mm DTA and the gamma index CL results fall within the TG 119 criteria, which is below 12.4. Conclusion: TG 119 methodology and recommendations have successfully been used to evaluate commissioning accuracy of VMAT. The CL value of the study could be used as a reference and recommendation to evaluate the accuracy and integrity of treatment planning and treatment delivery systems of VMAT and IMRT. Citation: Nainggolan A, Pawiro S. A. Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol. J Biomed Phys Eng. 2019;9(4):395-408. https://doi.org/10.31661/jbpe.v0i0.839.

Authors and Affiliations

S. A. Pawiro, A. Nainggolan

Keywords

Related Articles

Calculating Weighting Factors for Mixing Megavoltage Photon Beams to Achieve Desirable Dose Distribution in Radiotherapy

Background: In radiotherapy, low-energy photon beams are better adapted to the treated volume, and the use of high-energy beams can reduce hot spots in the radiation therapy. Therefore, mixing low and high energies with...

Simulation study of Hemodynamic in Bifurcations for Cerebral Arteriovenous Malformation using Electrical Analogy

Background and Objective: Cerebral Arteriovenous Malformation (CAVM) hemodynamic is disease condition, results changes in the flow and pressure level in cerebral blood vessels. Measuring flow and pressure without cathete...

Evaluation of Electron Specific Absorbed Fractions in Organs of Digimouse Voxel Phantom Using Monte Carlo Simulation Code FLUKA

Background: For preclinical evaluations of radiopharmaceuticals, most studies are carried out on mice. Electron specific absorbed fractions (SAF) values have had vital role in the assessment of absorbed dose. In past stu...

Assessment of Neutron Contamination Originating from the Presence of Wedge and Block in Photon Beam Radiotherapy

Background: One of the main causes of induction of secondary cancer in radiation therapy is neutron contamination received by patients during treatment. Objective: In the present study the impact of wedge and block on ne...

Closed- and Open-loop Deep Brain Stimulation: Methods, Challenges, Current and Future Aspects

Deep brain stimulation (DBS) is known as the most effective technique in the treatment of neurodegenerative diseases, especially Parkinson disease (PD) and epilepsy. Relative healing and effective control of disease symp...

Download PDF file
  • EP ID EP647171
  • DOI -
  • Views 97
  • Downloads 0

How To Cite

S. A. Pawiro, A. Nainggolan (2019). Dosimetric Evaluation of Volumetric Modulated Arc Therapy (VMAT) and Intensity Modulated Radiotherapy (IMRT) Using AAPM TG 119 Protocol. Journal of Biomedical Physics and Engineering, 9(4), 395-408. https://europub.co.uk/articles/-A-647171