Effect of Acquisition Parameters of SPECT Gamma Camera System on Image Quality

Journal Title: JOURNAL OF CONTEMPORARY MEDICINE - Year 2018, Vol 8, Issue 3

Abstract

OBJECTIVE: Quality control testing of a SPECT gamma camera is crucial in assessing the suitability of the camera for use in nuclear medicine department. The aim of the present study was to investigate the effect of gamma camera acquisition parameters on image quality.METHODS: Camera scanning was carried out using a double-headed gamma cameraon a total of 48 patients (29 female and 19 male, mean age: 47.4±11.1) referred to our department for myocardial perfusion scintigraphy and thyroid scintigraphy. Then, camera acquisition parameters were changed (for myocardial perfusion scintigraphy from 64x64 to 128x128 matrix change and for thyroid scintigraphy from 5 cm to 10 cm distance change), and scanning was repeated and images were analyzed.RESULTS: Left ventricle ejection fraction (EF) value in 64x64 matrix was calculated to be 62.7±8.8%. Smaller EF value of 48.9±10.3% was obtained in 128x128 matrix for the same patients. 99mTc-pertechnetate uptake percentage was 3.8±2.3% in measurements carried out at a distance of 10 cm. On the other hand, a higher uptake percentage of 6.2±3.6% was found for the same patients measured at a distance of 5 cm. CONCLUSION: In order to obtain proper imaging in SPECT gamma camera system, correct acquisition parameters should be used along with quality control tests for intrinsic flood-field uniformity and relative sensitivity.

Authors and Affiliations

Serdar Savaş Gül

Keywords

Related Articles

Deneysel epilepsi modelinde böbrekte bulunan aquaporin4 ve aquaporin2 kanallarının gen ekspresyonları

Epilepsinin temel oluşum mekanizmasında, beyinde bulunan çeşitli iyon kanallarının genlerindeki değişikliğin yer aldığı bilinmektedir. Potasyum kanallarındaki fonksiyonel değişikliklerin yanı sıra, bir su kanalı olan v...

Simulation Study on Performance of Balance Metrics in Propensity Score Weighting Method

Objective: In the situation that randomization is not avaliable, to minimize the biasness in treatment arm assignments, the use of propensity score weighting method and the assessment of performances related to results o...

Cervikal Ectopic Pregnancy: a case report

Cervical ectopic pregnancy which blastocyste  implants in the endocervical region is a rare form of ectopic pregnancy. Early diagnosis and treatment are very important due to prevention of mortality and preserving fertil...

Romatoid artritte plöropulmoner bulgular için risk faktörleri: Bir pilot çalışma

Amaç: Çalışmanın amacı, romatoid artritte (RA) pulmoner bulguların gelişiminde olası plöropulmoner bulguların paternini ve olası risk faktörlerini belirlemektir. Yöntem: Toplam 70 romatoid artrit hastası alındı. Akciğer...

Pulmoner Emboli Hastalarında Nötrofil/Lenfosit Oranı ve Trombosit/Lenfosit Oranı

Amaç: Çalışmanın amacı akut pulmoner embolisi olan hastalarda nötrofil/lenfosit oranının (NLR) ve plateled/lenfosit oranının (PLR) klinik önemini araştırmaktır.Yöntem: Ocak 2016 ile Aralık 2017 tarihleri ​​arasında akut...

Download PDF file
  • EP ID EP423523
  • DOI 10.16899/gopctd.411095
  • Views 100
  • Downloads 0

How To Cite

Serdar Savaş Gül (2018). Effect of Acquisition Parameters of SPECT Gamma Camera System on Image Quality. JOURNAL OF CONTEMPORARY MEDICINE, 8(3), 201-206. https://europub.co.uk/articles/-A-423523