Effect of Device Design on the Aerosolization of a Carrier-Based Dry Powder Inhaler—a Case Study on Aerolizer® Foradile®

Journal Title: The AAPS Journal - Year 2013, Vol 15, Issue 2

Abstract

The objective of this study is to investigate the effect of device design of the Aerolizer® on the aerosolization of a carrier-based dry powder inhaler formulation (Foradile®). The Aerolizer was modified by reducing the air inlet size and mouthpiece length to 1/3 of the original dimensions, or by increasing the grid voidage. Aerosolization of the powder formulation was assessed on a multi-stage liquid impinger at air flow rates of 30, 60, and 100 L/min. Coupled CFD-DEM simulations were performed to investigate the air flow pattern and particle impaction. There was no significant difference in the aerosolization behavior between the original and 1/3 mouthpiece length devices. Significant increases in FPF total and FPF emitted were demonstrated when the inlet size was reduced, and the results were explained by the increases in air velocity and turbulence from the CFD analysis. No significant differences were shown in FPF total and FPF emitted when the grid voidage was increased, but more drugs were found to deposit in induction port and to a lesser extent, the mouthpiece. This was supported by the CFD-DEM analysis which showed the particle–device collisions mainly occurred in the inhaler chamber, and the cross-grid design increased the particle–device collisions on both mouthpiece and induction port. The air inlet size and grid structure of the Aerolizer® were found to impact significantly on the aerosolization of the carrier-based powder.

Authors and Affiliations

Qi Tony Zhou, Zhenbo Tong, Patricia Tang, Mauro Citterio, Runyu Yang, Hak-Kim Chan

Keywords

Related Articles

Aggregation Kinetics for IgG1-Based Monoclonal Antibody Therapeutics

Monoclonal antibodies (mAbs) as a class of therapeutic molecules are finding an increasing demand in the biotechnology industry for the treatment of diseases like cancer and multiple sclerosis. A key challenge associated...

Efficacious Intestinal Permeation Enhancement Induced by the Sodium Salt of 10-undecylenic Acid, A Medium Chain Fatty Acid Derivative

The online version of this article (doi:10.1208/s12248-014-9634-3) contains supplementary material, which is available to authorized users.

Utilization of Gene Expression Signature for Quality Control of Traditional Chinese Medicine Formula Si-Wu-Tang

The online version of this article (doi:10.1208/s12248-013-9491-5) contains supplementary material, which is available to authorized users.

On the Incorrect Statistical Calculations of the Kinetica Software Package in Imbalanced Designs

This regulatory note supports the previous findings that suggest that the software package Kinetica, up to version 5.0.10, provides incorrect results for the 90% confidence intervals for the ratio test/reference where th...

Development of a Generic Anti-PEG Antibody Assay Using BioScale’s Acoustic Membrane MicroParticle Technology

Immunogenicity testing for PEGylated biotherapeutics should include methods to detect both anti-protein and anti-PEG antibodies (anti-PEG). Although some methods have been published for the detection of anti-PEG antibodi...

Download PDF file
  • EP ID EP681103
  • DOI  10.1208/s12248-013-9458-6
  • Views 58
  • Downloads 0

How To Cite

Qi Tony Zhou, Zhenbo Tong, Patricia Tang, Mauro Citterio, Runyu Yang, Hak-Kim Chan (2013). Effect of Device Design on the Aerosolization of a Carrier-Based Dry Powder Inhaler—a Case Study on Aerolizer® Foradile®. The AAPS Journal, 15(2), -. https://europub.co.uk/articles/-A-681103