Evaluation of dental X-ray apparatus in terms of patient exposure to ionizing radiation

Journal Title: Medycyna Pracy - Year 2017, Vol 68, Issue 4

Abstract

Background: The use of X-ray in dental procedures causes exposure of the patient to ionizing radiation. This exposure depends primarily on the parameters used in tooth examination. The aim of the study was to determine the patients exposure and to assess the technical condition of X-ray tubes. Material and Methods: Seventeen hundred dental offices were covered by the questionnaire survey and 740 questionnaires were sent back. Direct measurements were performed in 100 units by using the thermoluminescent detectors and X-ray films. Results: The results showed that the most commonly used exposure time is 0.22±0.16 s. The average entrance dose for the parameters used most commonly by dentists is 1.7±1.4 mGy. The average efficiency of X-ray tube estimated on the basis of exposures is 46.5±23.7 μGy/mAs. Conclusions: The study results indicate that the vast majority of X-ray tubes meet the requirements specified in the binding regulations. Med Pr 2017;67(4):491–496

Authors and Affiliations

Jerzy Olszewski, Małgorzata Wrzesień

Keywords

Related Articles

Challenges to occupational medicine in view of the problem of work-related diseases and the aging of workforce. Directions for further development and intentional changes in preventive care of employees in Poland

The system of occupational health care in Poland, based on occupational medicine service, takes care of almost 12.5 million employees subjected to over 4.5 million obligatory periodic medical check ups. This form of prov...

Management of tobacco smoking in the prison service units – The effects of the health promotion program

Background: In 2014 the health promotion program aimed at managing personnel smoking was implemented in the Polish prison service (PS) in cooperation with the Nofer Institute of Occupational Medicine (NIOM). It combined...

The usefulness of an optical monitor for the assessment of human exposure to fine dust in indoor air

Background: An important issue in the assessment of health risks related to air pollution with fine dust is the measurement of individual exposure. Such possibilities are provided by relatively easy to use optical monito...

Occupational diseases among healthcare and social workers in 2009–2016

Background: The aim of the paper is to present statistical data on the occurrence of occupational diseases among healthcare and social workers in Poland in 2009–2016. Material and Methods: All cards certifying that a cas...

Emission of nanometer size particles during selected processes with construction materials using

Background: The aim of the presented work was the assessment of occupational exposure to nanoparticles and ultrafine particles during selected processes of using construction materials. Material and Methods: The tests we...

Download PDF file
  • EP ID EP182815
  • DOI 10.13075/mp.5893.00506
  • Views 173
  • Downloads 0

How To Cite

Jerzy Olszewski, Małgorzata Wrzesień (2017). Evaluation of dental X-ray apparatus in terms of patient exposure to ionizing radiation. Medycyna Pracy, 68(4), 491-496. https://europub.co.uk/articles/-A-182815