Evaluation of Radiation Doses from Paediatric Chest X-ray Examination at Three Hospitals in South-West Nigeria

Journal Title: Physical Science International Journal - Year 2017, Vol 15, Issue 2

Abstract

The aim of this study is essentially to evaluate radiation doses to paediatric patients who were referred for chest X-ray examination between April, 2015 and December, 2016 in three hospitals South-West, Nigeria. The methodology employed in this study involved measurements based on the use of exposure parameters to determine Entrance Surface Dose (ESD). A total of one hundred (100) paediatric patients randomly chosen were included in this study. ESD was determined from exposure parameters using dose calculation software (Dosecal). The results of this study show that the range of Entrance Surface Doses (ESDs) obtained for 0-1year patient is 38.00-150.00 µGy, for 1-5 years is 43.34-194.00 µGy, for 5-10 years is 74.00-223.00 µGy and for 10-15 years is 87.60-292.00 µGy. The mean ESDs obtained in this study were found to be relatively lower than the ESDs obtained in other previous studies in Nigeria but higher than the recommended reference dose value given by the European guidelines. This indicates that more attention needs to be given to X-ray facilities in the Country and that X-ray departments need to review their practice in terms of training and re-training of radiographers, repeated chest X-ray procedures for paediatrics, calibration of X-ray machines and standardisation of processing chemicals in order to bring dose given to patients to optimum levels. This study has also revealed the need to have dedicated X-ray facilities for paediatrics throughout the country.

Authors and Affiliations

Osahon O. David, Bamidele Lateef

Keywords

Related Articles

Corrosion Inhibition of Mild Steel and Aluminium in 1 M Hydrochloric Acid by Leaves Extracts of Ficus sycomorus

Inhibitory effects of Ficus sycomorus leaves extracts on the corrosion of mild steel and aluminium in 1 M hydrochloric acid (HCl) solution was studied at temperature of 30°C using gravimetric technique. Corrosion rates of...

Design and Development of an Improved Palm Kernel Shelling Machine and Separator

A palm kernel processing (Cracking and Separating) unit was developed to crack effectively various sizes/diameters of palm kernel as well as to separate the palm kernel from the shell with the aim of reducing the challen...

Variability of Microwave Radio Refractivity and Field Strength over Some Selected Locations in Nigeria

In this study, results of the variation of radio field strength derived from the computation of surface refractivity through measurements of temperature, relative humidity and pressure across seven locations ((Akure (7.1...

A Simple Relative Humidity Sensor Employing Optical Fiber Coated with Lithium Chloride

A simple optical fiber relative humidity sensor was fabricated using a lithium chloride film coated on the distal end of sensing fiber. The sensing element, lithium chloride film whose refractive index is sensitive to mo...

Filter Design and Applications in Image Improvement

This work presents the performance analysis of different basic techniques used for the image restoration. Restoration is a process by which an image suffering from degradation can be recovered to its original form. Remov...

Download PDF file
  • EP ID EP321059
  • DOI 10.9734/PSIJ/2017/34269
  • Views 54
  • Downloads 0

How To Cite

Osahon O. David, Bamidele Lateef (2017). Evaluation of Radiation Doses from Paediatric Chest X-ray Examination at Three Hospitals in South-West Nigeria. Physical Science International Journal, 15(2), 1-6. https://europub.co.uk/articles/-A-321059