Evaluation of the Effect of Source Geometry on the Output of Miniature X-ray Tube for Electronic Brachytherapy through Simulation
Journal Title: Journal of Biomedical Physics and Engineering - Year 2018, Vol 8, Issue 1
Abstract
Objective: The use of miniature X-ray source in electronic brachytherapy is on the rise so there is an urgent need to acquire more knowledge on X-ray spectrum production and distribution by a dose. The aim of this research was to investigate the influence of target thickness and geometry at the source of miniature X-ray tube on tube output. Method: Five sources were simulated based on problems each with a specific geometric structure and conditions using MCNPX code. Tallies proportional to the output were used to calculate the results for the influence of source geometry on output. Results: The results of this work include the size of the optimal thickness of 5 miniature sources, energy spectrum of the sources per 50 kev and also the axial and transverse dose of simulated sources were calculated based on these thicknesses. The miniature source geometric was affected on the output x-ray tube. Conclusion: The result of this study demonstrates that hemispherical-conical, hemispherical and truncated-conical miniature sources were determined as the most suitable tools.
Authors and Affiliations
B. Barati, M. Zabihzadeh, M. J. Tahmasebi Birgani, N. Chegini, J. Fatahiasl, I. Mirr
A Glance at the Errors of Some Studies on the Health Effects of High Background Natural Radiation Areas
There is no place on the Earth, the planet we live on, where the natural background radiation level is zero. Since the birth and even in our fetal stage, we have been exposed to different sources of natural radiation. Li...
Simulation study of Hemodynamic in Bifurcations for Cerebral Arteriovenous Malformation using Electrical Analogy
Background and Objective: Cerebral Arteriovenous Malformation (CAVM) hemodynamic is disease condition, results changes in the flow and pressure level in cerebral blood vessels. Measuring flow and pressure without cathete...
Alzheimer ’s Disease: Possible Mechanisms Behind Neurohormesis Induced by Exposure to Low Doses of Ionizing Radiation
In 2016, scientists reported that human exposure to low doses of ionizing radiation (CT scans of the brain) might relieve symptoms of both Alzheimer’s disease (AD) and Parkinson disease (PD). The findings were unbelievab...
Dosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography
Background: While the benefits of cone-beam computed tomography (CBCT) are well known in maxillofacial imaging, the use of this modality is not risk-free. Objective: The aim of this study was to evaluate the exposure dos...
Effects of Laser Physical Parameters on Lesion Size in Retinal Photocoagulation Surgery: Clinical OCT and Experimental Study
Introduction: The aim of the present study was to determine burn intensity in retinal laser photocoagulation based on laser parameters; wavelength, power, beam size and pulse duration, using Optical Coherence Tomography...