HbA2 ölçümü için BioRad D-10TM ve Tosoh HLC 723 G8 HPLC sistemlerinin karşılaştırılması (Comparison of the BioRad D-10TM and Tosoh HLC 723 G8 HPLC instruments for the detection of HbA2)
Journal Title: Turk Hijyen ve Deneysel Biyoloji Dergisi - Year 2016, Vol 73, Issue 3
Abstract
Amaç: Özellikle Akdeniz bölgesinde önemli bir halk sağlığı sorunu olan hemoglobin bozuklukları taramasında hemoglobin sub tiplerini ayırt etmek amacıyla en yaygın kullanılan yöntem HPLC (High Performance Liquid Chromatography)’dir. Pek çok marka ve model HPLC sistemleri ile hemoglobinopati taraması yapılmakta ancak sistemlerin birbirleri ile uyumu bilinmemektedir. Bu çalışmanın amacı; D-10TM Hemoglobin testingsystem (BioRad,Hercules,USA) marka HPLC cihazı ile Tosoh HLC 723 G8 (TosohBioscience, Japan) HPLC cihazını hemoglobinopati tanı ve taraması açısından önemli olan HbA2 parametresi ölçümü bakımından karşılaştırılmasıdır. Yöntemler: Çalışmaya Antalya Eğitim ve Araştırma Hastanesi Merkez Laboratuvarı Biyokimya Bölümüne 1-31 Ocak 2016 tarihleri arasında talasemi taraması için başvuran 12’si erkek, 18’i kadın toplam 30 kişi alındı. Her iki cihazda tek seferde çalışılan hemoglobin A2 (HbA2) değerlerinin farkları için bağımlı iki grup ve iki ölçüm fark testleri, istatistiksel uyum ve ilişki analizi için iki-yönlü rastgele sınıf içi korelasyon katsayısı (SKK), Cohen’s kappa katsayısı, korelasyon katsayısı (r) ve Bland Altman Analizi yapıldı. Sonuçlar uygun grafiklerle sunuldu. Bulgular: Analizler sonucunda Tosoh HLC 723 G8 cihazında örneklerin 14’ü (%46.66) ve BioRad D-10TM cihazında örneklerin dokuz’u (%30) β -talasemi taşıyıcısı olarak saptandı. Her iki cihazda ölçülen HbA2 değerleri arasında anlamlı fark saptandı (z= -4.487; p <0.001). HbA2 değerleri için sınıf içi korelasyon katsayısı 0.898, uyum katsayısı 0.783, Spearman korelasyon katsayısı 0.944, Cohen’s Kappa değeri 0.658 ve Bland Altman Analizi sonucu yanlılık değeri 0.75 (Alt Limit: -0.49; Üst Limit: 1.99) olarak bulundu. Sonuç: Korelasyon, regresyon ve diğer istatistiksel analizlere göre her iki cihaz arasındaki uyum yetersiz olarak bulundu. Objective: HPLC (High Performance Liquid Chromatography) is the most commonly used method to differentiate the subtypes of hemoglobin in screening for hemoglobinopathy which is a major public health care problem especially in Mediterranean region. Today, different models and trademarks of HPLC equipment have been used in screening for hemoglobinopathy. However, the agreement of the results of the analyses performed by these types of equipment are unknown. In this study, the aim was to investigate the agreement between the results of the analyses of HbA2 values with D-10TM Hemoglobin testing system (BioRad,Hercules,USA) and Tosoh HLC 723 G8 (TosohBioscience, Japan) trademark. Methods: A total of 30 individuals, 12 males and 18 females, applied to the Antalya Training and Research Hospital laboratory between January 1 and January 31, 2016 for thalassemia screening were included to this study. Statistical correlations of hemoglobin A2 (HbA2) values, analyzed in each analyzer as a single analysis from each sample, were evaluated using two-way random intraclass correlation coefficient (ICC), Cohen’s kappa coefficient, correlation coefficient (r) and Bland Altman analysis. Results are presented with graphics. Results: As a result of the analysis, 14 of the samples (46.66%) were detected as – thalassemia carrier with Tosoh HLC 723 G8 system and nine of the samples (30%) were detected as – thalassemia carrier with BioRad D-10TM system. There was significant differance in HbA2 values measured with both systems (z= -4.487; p <0.001). For HbA2 values; intraclass correlation coefficient was 0.898, coefficeint of concordance was 0.783, Spearman corelation coefficiency was 0.944, Cohen’s Kappa value 0.658 and Bland Altman analysis bias value was 0.75 (LowerLimit: -0.49; Upper Limit: 1.99). Conclusion: According to the regression, correlation and other statistical analyses, agreement between the results of HbA2 values measured by the two analyzers were found to be insufficient.
Authors and Affiliations
Güzin AYKAL, Fazıla ATAKAN-ERKAL, Ayşenur YEĞİN, Esin EREN, Necat YILMAZ
Polimerik veziküller ve biyolojik uygulamaları (Polymeric vesicles and biological applications)
Bilimin gelişmesindeki temel yaklaşım olan doğayı taklit etme çabasının günümüzdeki en son örneklerinden birisi polimerik veziküler sistemlerdir. Hücresel bölümlendirmenin yapıtaşı olan fosfolipitlerin kendiliğinden oluş...
Kuaför salonlarındaki kimyasallara mesleki maruziyet ve sağlık riski (Occupational exposure to the chemicals in hairdressing salons and health risk)
Kozmetik ürünlerinin yapımında, 5.000’in üstünde kimyasal kullanımı söz konusudur. Kuaförlük mesleği çalışanları, başlıca saç ürünleri ve yanı sıra tırnak ve cilt bakımında kullanılan çok sayıda kozmetik ürüne mesleki ol...
Asymptomatic bacteriuria, urinary tract infection and risk factors in women with type 2 diabetes mellitus and impaired glucose tolerance (Tip 2 Diyabet hastalığı ve bozulmuş glikoz toleransı olan kadınlarda asemptomatik bakteriüri ve üriner sistem enfeksiyonları)
Objective: We aimed to compare a type 2 diabetic women groups with a women group with impaired glucose tolerance (IGT) for the presence of urinary tract infection (UTI) and/or asymptomatic bacteriuria (ASB) and related r...
Olgu sunumu; Yanıkta izole edilen Bacillus licheniformis (Case report: The burn isolated Bacillus licheniformis)
ÖZET Bacillus licheniformis doğada yaygın olarak bulunabilen, saprofitik bir bakteri olup insanlarda bakteriyemi, peritonit, gıda zehirlenmeleri gibi hastalıklara neden olabilmektedir. Medikal olarak kullanılan basitrasi...
Tick attachment sites in humans living in the Tokat province of Turkey (Tokat ilinde yaşayan insanlardaki kene tutunma bölgelerinin değerlendirilmesi)
Objective: The aim of this study was to evaluate the attachment sites of ticks and the demographic properties of patient infested by ticks in Tokat province, a Crimean-Congo hemorrhagic fever (CCHF) endemic region in Tur...