HPLC method for simultaneous determination of oxytocin and clorobutanol in injectable solutions
Journal Title: Medicamentul Veterinar / Veterinary Drug - Year 2013, Vol 7, Issue 2
Abstract
The objective of this work was to establish and validate a HPLC method with UV detection for simultaneous determination of oxytocin and clorobutanol in veterinary injectable formulations. The method is based on European Pharmacopoeia monograph for oxytocin concentrated solution. Chromatographic separation was achieved on a Syn Chropack RP100 C18 column (250 x 4,6mm, 5 μm) with a mobile phase consisting of solutia A: sodium dihydrogen phosphate buffer 0,13 M and solution B: acetonitrile-water (1:1, v/v) with gradient elution (30 % B for 1 min, 30 % B to 60 % B in 30 min, return to initial concentration and echilibration for 15 min before the following injection), at a flowrate of 1 ml/min and detection at 220 nm. The retention times for oxytocin and clorobutanol were about 15 min and 26 min respectively. The degradation products of oxytocin eluted at retention times smaller than 14 min. The resolution between oxytocin and the nearest impurity fulfilled the USP monograph requirements of at least 1,5. The linearity of the method has been settled from 2,5 UI/ml to 20 UI/ml for oxytocin and 1,25 mg/ml to 10 mg/ml for clorobutanol. In these ranges the corelation coefficients were higher than 0,9950. The method allows the separation of oxytocin from degradation products and clorobutanol and could be applicable to quality control of injectable products containing oxytocin and clorobutanol. The method was validated in terms of selectivity, linearity, precision and accuracy.
Authors and Affiliations
Ana Csuma, Despina Niţă, Luciana Paraschiv
Development and validation of a RP–HPLC method for the quantization studies of albendazole suspensions dosage forms of Rombendazol
An isocratic high–performance liquid chromatography (HPLC) procedure was developed for quantitative determination of albendazole in suspensions dosage forms of ROMBENDAZOL. HPLC separation was carried out by reversed pha...
Earthworms as biomarkers for detecting soil pollution around swine farms in Timiș County
By using several samples collected from different farms distances it can be determined the degree of farms pollution and their effects on ecosystems. In order to avoid any error the soil samples were tested in comparison...
Dezvoltarea si validarea metodei RP- HPLC de determinare cantitativă a furazolidonei si metronidazol din produsul Enteroguard M
A fost dezvoltată si validată o metodă isocratică de lichid cromatografie de performantă înaltă pentru determinarea cantitativă a metronidazolului si furazolidonei din comprimate Enteroguard M. Separarea HPLC a fost real...
Juvaenes Alumni
În cele ce urmează sunt prezentate două doctorande aparținătoare Școlii Doctorale de la Facultatea de Medicină Veterinară din Timișoara admise la studiile doctorale în acest an.
Metoda HPLC pentru determinarea simultana a oxitocinei şi clorbutanolului din solutii injectabile
Obiectivul lucrării a fost stabilirea şi validarea unei metode HPLC cu detecţie în ultraviolet (UV) pentru determinarea simultană a oxitocinei şi clorbutanolului din produse medicinale veterinare injectabile. Metoda se b...