Interchangeability of corneal curvature and asphericity measurements provided by three different devices

Journal Title: International Journal of Ophthalmology - Year 2019, Vol 12, Issue 3

Abstract

"AIM: To evaluate the interchangeability of keratometric and asphericity measurements provided by three measurement systems based on different optical principles. METHODS: A total of 40 eyes of 40 patients with a mean age of 34.1y were included. In all cases, a corneal curvature analysis was performed with IOL-Master (IOLM), iDesign 2 (ID2), and Sirius systems (SIR). Differences between instruments for flattest (K1) and steepest (K2) keratometric readings, as well as for magnitude and axis of corneal astigmatism were analyzed. Likewise, differences in asphericity (Q) between SIR and ID2 were also evaluated. RESULTS: Mean differences between devices for K1 were 0.20±0.21 (P<0.001), -0.12±0.36 (P=0.046) and -0.32±0.36 D (P<0.001) for the comparisons IOLM-SIR, IOLM-ID2 and SIR-ID2, respectively. The ranges of agreement for these comparisons between instruments were 0.41, 0.70, and 0.70 D. For K2, mean differences were 0.31±0.33 (P<0.001), -0.08±0.43 (P=0.265) and -0.39±0.38 D (P<0.001), with ranges of agreement of 0.65, 0.84, and 0.74 D. Concerning magnitude of astigmatism, ranges of agreement were in the limit of clinical relevance (0.49 D, P=0.011; 0.55 D, P=0.386; 0.43 D, P=0.05). In contrast, ranges of agreement were clinically relevant for astigmatic axis (26.68º, 33.83º and 18.37º, P≥0.121) and for Q between SIR and ID2 (0.16, P<0.001). CONCLUSION: The keratometric corneal power, astigmatic axis and asphericity measurements provide by the three systems evaluated cannot be considered as interchangeable, whereas measurements of corneal astigmatism obtained with SIR and ID2 can be considered as interchangeable for clinical purposes."

Authors and Affiliations

David P. Piñero

Keywords

Related Articles

Analysis of proteomic differences between liquefied after-cataracts and normal lenses using two-dimensional gel electrophoresis and mass spectrometry

"AIM: To analyze and identify the proteomic differences between liquefied after-cataracts and normal lenses by means of liquefied chromatography-tandem mass spectrometry (LC-MS/MS). METHODS: Three normal lenses and three...

Prevalence and treatment patterns of ranibizumab and photodynamic therapy in a tertiary care setting in Malaysia

"AIM: To describe the prevalence and changes in treatment patterns of ranibizumab and photodynamic therapy (PDT) among retinal disease patients who attended the Ophthalmology Clinic in the tertiary care Hospital Selayang...

Human eye ocular component analysis for refractive state and refractive surgery

"AIM: To analyze the clinical factors influencing the human vision corrections via the changing of ocular components of human eye in various applications; and to analyze refractive state via a new effective axial length....

Modified deep sclerectomy combined with Ex-PRESS filtration device versus trabeculectomy for primary open angle glaucoma

"AIM: To compare the efficacy of modified deep sclerectomy combined with Ex-PRESS shunt versus trabeculectomy in primary open angle glaucoma. METHODS: This is a prospective cohort comparative single-center study. Forty-n...

Download PDF file
  • EP ID EP591471
  • DOI 10.18240/ijo.2019.03.10
  • Views 50
  • Downloads 0

How To Cite

David P. Piñero (2019). Interchangeability of corneal curvature and asphericity measurements provided by three different devices. International Journal of Ophthalmology, 12(3), -. https://europub.co.uk/articles/-A-591471