Longitudinal Stent Deformation at Aneurysm Site: Flexibility at the Expense of Longitudinal Integrity
Journal Title: Journal of Clinical and Diagnostic Research - Year 2017, Vol 11, Issue 7
Abstract
Longitudinal stent deformation is a recently described complication of percutaneous coronary intervention. Novel stents with thin struts and reduced number of fixed links between cells improve flexibility and deliverability but in certain cases it may reduce longitudinal strength and thereby increase the risk of longitudinal deformation. Although longitudinal deformation of coronary stents is an infrequent finding, it requires clinical attention as it may lead to catastrophic clinical outcomes. We report a case of longitudinal deformation of coronary stent observed at our institution while treating ostial lesion and aneurysm of left anterior descending artery. Longitudinal deformation was identified during the procedure and treated with the deployment of another stent. Three-month follow-up of the patient was found satisfactory without any incidence of stent thrombosis.
Authors and Affiliations
Bhupesh R Shah
Assessment of Disease Severity and Role of Cytomegalo Virus Infection in Patients with Ulcerative Colitis
Introduction: Course of Ulcerative Colitis is characterized by intermittent flares interposed between variable periods of remission. Identification of exacerbating factors and appropriate assessment of disease activity a...
Comparison of Magnesium Sulphate Protocols by Weight-Adjusted Versus Two Grams Per Hour for Preventing Convulsion in Preeclampsia: A Randomised Controlled Trial
ABSTRACT Introduction: Magnesium Sulphate (MgSO4) is the most effective drug used for seizure prevention in preeclampsia; however, the dosage of MgSO4 is still in controversy. Maternal weight is an important factor for t...
Comparison of the Effect of PRP, PRF and Induced Bleeding in the Revascularization of Teeth with Necrotic Pulp and Open Apex: A Triple Blind Randomized Clinical Trial
Introduction: Treatment of a tooth with necrotic pulp and open apex is a special challenge to the clinicians. Apexification with calcium hydroxide and MTA barrier technique fails to induce continued root maturation which...
All Mycobacteria are not Tubercular: A Case Report
Abstract Mycobacterium tuberculosis is a common infection in developing countries with high morbidity and mortality. However, Non-Tuberculous Mycobacteria (NTM) are common environmental organisms which closely resemble t...
Occlusal Traits of Primary Dentition among Pre-School Children of Mehsana District, North Gujarat, India
Introduction: Aesthetic and functional problems in the dentofacial complex among children are often attributed to the disturbances in the dentofacial growth and development. This can be recognized during routine dental v...