Measurements of HbA1c by High Performance Liquid Chromatography in D-10 analyzer and Immunological Method by Beckman Coulter AU480 System: A Comparative Study

Journal Title: Journal of Enam Meical College - Year 2012, Vol 2, Issue 2

Abstract

Background: HbA1c can be measured by different methods. Enzymatic, boronate affinity chromatographic and cation-exchange high performance liquid chromatographic (HPLC) methods are considered as gold standard methods. To make the test cost-effective many laboratories use the immunological method installed in a chemistry analyzer. Objective: To compare the values of HbA1c measured by high performance liquid chromatographic method and immunological method in two laboratory settings. Materials and Methods: This observational and comparative study was conducted in Bangladesh Institute of Health Sciences (BIHS) and Enam Medical College, Savar, Dhaka. HbA1c levels measured by HPLC in BIHS were compared with that measured by immunological method in Enam Medical College, Savar, Dhaka. p values <0.05 were considered significant. Results: Statistically no difference was observed in mean HbA1c% between two methods in <7.0% HbA1c group and 7.1-10.0% HbA1c group. Significant statistical difference was found in mean HbA1c >10.0% group, but it was within acceptable limit. Conclusion: HbA1c can be measured by immunological method installed in an automated chemistry analyzer to make the test costeffective. But HPLC method is preferable.

Authors and Affiliations

Md Aminul Haque Khan, Mst Rokshana Rabeya, Muhammad Saiedullah

Keywords

Related Articles

Comparative Study between Slow Shock Wave Lithotripsy and Fast Shock Wave Lithotripsy in the Management of Renal Stone

Background: Renal calculi are frequent causes of ureteric colic. Extracorporeal shock wave lithotripsy is the most common treatment of these stones. It uses focused sound waves to break up stones externally. Objective:...

Significance of Immunohistochemistry in Accurate Characterisation of Hodgkin Lymphoma

Background: Hodgkin lymphomas are malignant disorders of cells residing in lymphoid tissue and containing Reed-Sternberg (RS) cells and its variants and account for about 0.7% of all new cancers. Hodgkin lymphomas are...

Histopathological Evaluation of Lymph Node Biopsies: A Hospital Based Study

Background: Lymphadenopathy is a common manifestation of a large variety of disorders, both benign and malignant. It is essential to define the pattern of disorders presenting primarily as lymph node enlargement in a p...

Fasciolopsis buski Infection in a 50-Year-Old Female

Fasciolopsiasis is a disease caused by the largest food-borne intestinal trematode known as Fasciolopsis buski. Here we report a case of a 50-year-old female who presented with upper abdominal pain and vomiting for one...

Evaluation of Recently Developed Regression Equation with Direct Measurement of Low-density Lipoprotein Cholesterol in a Bangladeshi Population

Background: Meaningful underestimation of low-density lipoprotein (LDL) cholesterol is an important shortcoming of Friedewald’s formula (FF) at higher triglyceride (TG) levels. Recently a regression equation (RE) has b...

Download PDF file
  • EP ID EP36148
  • DOI https://doi.org/10.3329/jemc.v2i2.12839
  • Views 323
  • Downloads 0

How To Cite

Md Aminul Haque Khan, Mst Rokshana Rabeya, Muhammad Saiedullah (2012). Measurements of HbA1c by High Performance Liquid Chromatography in D-10 analyzer and Immunological Method by Beckman Coulter AU480 System: A Comparative Study. Journal of Enam Meical College, 2(2), -. https://europub.co.uk/articles/-A-36148