Mechanical Fatigue Analysis Comparing Two Locking Plates in a Metaphyseal Fracture Model of the Distal Ulna

Journal Title: Journal of Musculoskeletal Disorders and Treatment - Year 2016, Vol 2, Issue 3

Abstract

The purpose of the study was to test the mechanical fatigue properties (fatigue limit and fatigue strength) of two locking plates designed for the distal ulna. Distal Ulna Plate 2.5 (Medartis, Basel, Switzerland) and 2.0 mm LCP Distal Ulna Plate (DePuySynthes, West Chester (PA), USA) were used. A metaphyseal fracture situation was chosen (AO 23-A1.3). Six plate-screw constructs of each design were mounted on custom made fixtures using a Sawbone's 3D ulna model. Testing was performed on monoaxial material testing machines. Endpoints were fatigue strength (load) and fatigue limit (cycles) when the sample fails. Initial load was set at Fmax 10 N. Sinusoidal loading was carried out at 4 Hz. After 50'000 cycles load was increased by 15% and again by 15% after each additional 10'000 cycles. 2.5 mm plates showed median fatigue strength of 40.5 N (range 5.3) and median fatigue limit of 141'307 cycles (range 5'436). 2.0 mm plates 13.2 N (range 5.2) and 67'287 cycles (range 28'718) respectively. Distal Ulna Plate 2.5 showed better mechanical properties.

Authors and Affiliations

Keywords

Related Articles

Survey of Pain Therapies in Marfan Syndrome

Objective Pain in Marfan syndrome is common, although frequently under diagnosed and undertreated. Few studies have investigated the treatment of pain symptoms in Marfan syndrome and no study has reported on the use of o...

Outcomes of Non-operative Management of Deep Gluteal Syndrome - A Case Series of Six Patients

Introduction:An increased understanding of posterior hip anatomy and biomechanics has led to advances in the diagnosis and management of Deep Gluteal Syndrome (DGS). Currently, there is limited information detailing the...

Minimally Invasive Percutaneous Screw Fixation of Phalangeal Fractures of the Hand

Aim This study present the results of an effective, minimally invasive treatment method for displaced diaphyseal and articular fractures of hand phalanges, which are difficult to manage and frequently involve complicatio...

Is Bony Evidence of Enthesial Reaction Sufficient for Differential Diagnosis?

Background Enthesial reaction, as a stress or disease marker, has been a generalized perspective, largely untested as to its veracity. Perhaps valid with soft tissue visualization by computerized tomography or magnetic r...

Perspectives on Improving the Efficacy of PRP Treatment for Tendinopathy

Platelet-rich plasma (PRP), popularly called PRP, is almost like a "house-hold" name these days because of its wide recognition by physicians and patients due to its prevalent use. The popularity of PRP is due to its eff...

Download PDF file
  • EP ID EP348042
  • DOI 10.23937/2572-3243.1510018
  • Views 108
  • Downloads 0

How To Cite

(2016). Mechanical Fatigue Analysis Comparing Two Locking Plates in a Metaphyseal Fracture Model of the Distal Ulna. Journal of Musculoskeletal Disorders and Treatment, 2(3), 1-4. https://europub.co.uk/articles/-A-348042