MTS-6 detectors calibration by using 239Pu-Be neutron source

Journal Title: Medycyna Pracy - Year 2017, Vol 68, Issue 6

Abstract

Background: Thermoluminescent detectors, type MTS-6, containing isotope 6Li (lithium) are sensitive in the range of thermal neutron energy; the 239Pu-Be (plutonium-and-beryllium) source emits neutrons in the energy range from 1 to 11 MeV. These seemingly contradictory elements may be combined by using the paraffin moderator, a determined density of thermal neutrons in the paraffin block and a conversion coefficient neutron flux to kerma, not forgetting the simultaneous registration of the photon radiation inseparable from the companion neutron radiation. The main aim of this work is to present the idea of calibration of thermoluminescent detectors that consist of a 6Li isotope, by using 239Pu-Be neutron radiation source. Material and Methods: In this work, MTS-6 and MTS-7 thermoluminescent detectors and a plutonium-and-beryllium (239Pu-Be) neutron source were used. Paraffin wax fills the block, acting as a moderator. The calibration idea was based on the determination of dose equivalent rate based on the average kerma rate calculated taking into account the empirically determined function describing the density of thermal neutron flux in the paraffin block and a conversion coefficient neutron flux to kerma. Results: The calculated value of the thermal neutron flux density was 1817.5 neutrons/cm2/s and the average value of kerma rate determined on this basis amounted to 244 μGy/h, and the dose equivalent rate 610 μSv/h. The calculated value allowed for the assessment of the length of time of exposure of the detectors directly in the paraffin block. Conclusions: The calibration coefficient for the used batch of detectors is (6.80±0.42)×10–7 Sv/impulse. Med Pr 2017;68(6):705–710

Authors and Affiliations

Małgorzata Wrzesień, Łukasz Albiniak, Hiba Al-Hameed

Keywords

Related Articles

OCENA NARAŻENIA NA PROMIENIOWANIE JONIZUJĄCE TECHNIKÓW OBSŁUGUJĄCYCH GAMMAKAMERY

Wstęp: Ochrona personelu placówki medycyny nuklearnej przed zagrożeniami wynikającymi ze stosowania promieniowania jonizującego jest istotnym aspektem ochrony radiologicznej osób fizycznych. Celem pracy była oceny naraż...

Health and quality of life vs. occupational activity

The level of quality of life and health status of the population largely depends on the determinants related to occupational activity. The results of reviewed bibliography indicate a significant and growing importance of...

Spine pain in the firefighter profession

Background: Musculoskeletal disorders mainly affecting the spine constitute one of the major social problems. Frequently, they constitute the cause of prolonged absence at work. Working in a fire brigade is specific beca...

Assessment of psychophysical capacities for professional work in late middle age and at the beginning of old age

Background: In Poland average life expectancy extends, while the number of the elderly who are active in the labor market decreases. Material and Methods: The study population consisted of 429 people aged 50–70 years old...

Optimization of ultra-performance liquid chromatography (UPLC) with fluorescence detector (FLD) method for the quantitative determination of selected neurotransmitters in rat brain

Background: Glutamate (Glu) and γ-aminobutyric acid (GABA) are the main neurotransmitters in the central nervous system for excitatory and inhibitory processes, respectively. Monitoring these neurotransmitters is an esse...

Download PDF file
  • EP ID EP228946
  • DOI 10.13075/mp.5893.00650
  • Views 159
  • Downloads 0

How To Cite

Małgorzata Wrzesień, Łukasz Albiniak, Hiba Al-Hameed (2017). MTS-6 detectors calibration by using 239Pu-Be neutron source. Medycyna Pracy, 68(6), 705-710. https://europub.co.uk/articles/-A-228946