Optimization of intraocular lens constant of a hybrid intraocular lens

Journal Title: Eye Science (Yanke Xuebao) - Year 2023, Vol 38, Issue 12

Abstract

[Objective:] To assess the benefits of intraocular lens (IOL) constant optimization of Alcon Acrysof IQ PanOptix TFNT00 (PanOptix) on the accuracy of IOL power calculation, and the effects of constant optimization between different axial length (AL) groups were further compared. [Methods:] Patients who underwent phacoemulsification and implantation with PanOptix IOL between June, 2021 and March, 2022 were included in this retrospective study. The preoperative biological ocular parameters, implanted IOL power, and subjective 1-3 months postoperative refraction were collected. Combined with SRK/T, Hoffer Q, Holladay 1 and Haigis formulas, the optimized IOL constant A, surgeon factor (SF), post-surgery anterior chamber depth (pACD), and a0, a1, a2 were back-calculated. Refractive outcomes using optimized IOL constants were re-calculated combined with the corresponding formulas. Compare the mean absolute error (MAE), medium absolute error (MedAE) and percentage of eyes with IOL prediction errors (PE) within ±0.25, ±0.50, ±0.75 and ±1.00 (diopter)D when using the optimized constants and the manufacture constants. Patients were divided into two groups according to AL (non-high myopia: <26.0 mm; high myopia: ≥26 mm), compare the difference of IOL constant optimization between AL subgroups. [Results:] A total of 92 eyes of 54 patients were enrolled. The manufacture lens constant of A, pACD, SF, a0, a1 and a2 are respectively 119.1, 5.63, 1.83, 1.39, 0.4 and 0.1; and the optimized values are respectively 119.35, 6.14, 2.36, −3.42, 0.12 and 0.34. In all patients group, with manufacture lens constant, the MAE values of SRKT, Hoffer Q, Holladay 1 and Haigis formula are 0.44, 0.50, 0.54, 0.46 D; with optimized lens constants, the MAE values are 0.43, 0.54, 0.51, 0.35 D, and there is a statistical difference of Haigis formula after optimization (P=0.001). In non-high myopia group, with manufacture lens constant, the MAE values are 0.46, 0.40, 0.40, 0.42 D; with optimized lens constants, the MAE values are 0.46, 0.38, 0.39, 0.38 D, and no statistical difference has been found (P>0.05). In high myopia group, with manufacture lens constant, the MAE values are 0.42, 0.59, 0.66, 0.50 D; with optimized lens constants, the MAE values are 0.36, 0.48, 0.47, 0.31 D, and there are statistical differences of Holladay 1 and Haigis formula after optimization (P=0.020, 0.002). [Conclusion:] IOL constant optimization of PanOptix IOL can improve the accuracy of IOL calculation, which is more significant in the high myopia group.

Authors and Affiliations

Suowang ZHOU, Abulimiti ADILAMU, Yongchang TAN, Jiying SHEN, Jin YANG, Xu CHEN

Keywords

Related Articles

Artificial intelligence advances a new model of hierarchic diagnosis and treatment for Cataract

With the rapid development of artificial intelligence (AI) technology, its application in the field of healthcare is bringing revolutionary changes. Cataracts, as one of the most common reversible visual impairments worl...

A seat for position nursing after pars plana vitrectomy in an ophthalmic day ward

To meet the needs of the postoperative position nursing after pars plana vitrectomy in the day surgery mode, this research team has designed a nursing chair [patent number: National Uutility Model Patent (ZL 2017 2 09424...

Efficacy of auricular points plaster therapy on preventing gastrointestinal reaction caused by fundus fluorescein angiography

[Objective:] To explore the prevention efficacy of auricular points plaster therapy on gastrointestinal reaction caused by fundus fluorescein angiography (FFA). [Methods:] We selected 583 patients who underwent fundus fl...

Epithelial downgrowth: a clinicopathologic report of a case of secondary glaucoma

Epithelial downgrowth is a rare yet serious complication after ocular trauma or anteriorsegmental surgery. It can lead to decompensation of corneal endothelium, secondary glaucoma or other serious complications, among wh...

Tuberous sclerosis complex presenting as retinal hamartomas in a preterm infant: A case report

Tuberous sclerosis complex is a multisystemic disease with an autosomal dominant inheritance pattern. Missed diagnosis and misdiagnosis are common for patients with single manifestation in the early stage. There are few...

Download PDF file
  • EP ID EP764706
  • DOI 10.12419/2308230001
  • Views 16
  • Downloads 0

How To Cite

Suowang ZHOU, Abulimiti ADILAMU, Yongchang TAN, Jiying SHEN, Jin YANG, Xu CHEN (2023). Optimization of intraocular lens constant of a hybrid intraocular lens. Eye Science (Yanke Xuebao), 38(12), -. https://europub.co.uk/articles/-A-764706