Sysmex XN1000 versus Manual Method in Leukopenic Blood Samples

Journal Title: Journal of Clinical and Diagnostic Research - Year 2018, Vol 12, Issue 4

Abstract

ABSTRACT Introduction: Complete Blood Cell (CBC) and differential leukocyte tests are one of the most common haematology laboratory investigations. Immature Granulocytes (IGs) form a part of extended leukocyte differential count and is considered as a promising tool for screening and early detection of bacteraemia. Differential leukocyte analysis provides information regarding the underlying pathologic process. Manual method of performing these counts is labor-intensive, time-consuming especially in leukopenic blood samples, lacks reproducibility and objectivity and therefore suffers from increased inter observer variation. Automated counts provided by the haematology analysers on the other hand are supposed to be more precise, objective and help to reduce the Turnaround Time (TAT). Aim: Comparison of Differential Leukocyte Count (DLC) and IG flagging by automated haematology analyser XN1000 with findings of manual method in leukopenic blood samples. Materials and Methods: The study included 346 patients considering the inclusion and exclusion criteria. CBC analysis was done using Sysmex XN1000. All leukopenic samples enumerated by the analyser were included and peripheral smears were made for each of them. Manual counts were correlated with cell counter counts. Statistical analysis was done using linear regression and correlation analysis. Mean counts were compared using Student's t-test. Pearson’s correlation coefficient and regression analysis was used for correlation studies. Bland-Altman plot analysis was also performed to compare the two measurement techniques. Results: Correlation coefficient for neutrophils, lymphocytes, eosinophils, monocytes and basophils proportional counts were 0.998, 0.992, 0.996, 0.771 and 0.570 respectively; and for absolute counts 0.999, 0.994, 0.996, 0.825 and 0.585 respectively. IG proportional and absolute counts correlation coefficients were 0.898 and 0.908 respectively. Sensitivity was 97.30%, specificity 91.91% and efficiency 92.49% for IG flagging. Conclusion: Strong positive correlation was observed between automated and manual neutrophil, lymphocyte, eosinophil, monocyte and IGs proportional and absolute counts. Basophil showed moderate degree of correlation for both proportional and absolute counts. IG flagging by Sysmex XN1000 showed high sensitivity, specificity and efficiency.Thus, automation seems to be a suitable, reliable and practical option for leukopenic blood samples keeping in view the results obtained by this analyser.

Authors and Affiliations

Divya Pursnani, Surekha B Hippargi

Keywords

Related Articles

Autoimmune Haemolytic Anaemia-A Spectrum of Presentation in Children

Autoimmune Haemolytic Anaemias (AIHAs) are rare in children. They can be either a primary disease or secondary to/triggered by a host of other clinical conditions. We present five interesting cases of paediatric AIHA ass...

Successful Outcome with Transradial Primary Percutaneous Coronary Intervention During Cardiopulmonary Resuscitation

Transradial access for ST Elevation Myocardial Infarction (STEMI) has been shown to reduce mortality, major adverse cardiac events and bleeding, when compared with femoral access. Often patients with STEMIs present in th...

A Study on Chronic Proton Pump Inhibitor Usage and Small Intestinal Bacterial Overgrowth (SIBO)

ABSTRACT Introduction: Small Intestinal Bacterial Overgrowth (SIBO) has a wide clinical spectrum, and may be associated with both maldigestion and malabsorption. Proton Pump Inhibitors (PPI) are one of the most widely pr...

Molecular Diversity of Hepatitis B Virus X Gene in Central Kerala, Southern India

ABSTRACT Introduction: Hepatitis B Virus (HBV) causes diverse clinical manifestations ranging from acute fulminant Hepatitis, Chronic Hepatitis (CHB) to liver cirrhosis and Haepatocellular Carcinoma (HCC). The HBV X gene...

Download PDF file
  • EP ID EP522525
  • DOI 10.7860/JCDR/2018/32516.11447
  • Views 75
  • Downloads 0

How To Cite

Divya Pursnani, Surekha B Hippargi (2018). Sysmex XN1000 versus Manual Method in Leukopenic Blood Samples. Journal of Clinical and Diagnostic Research, 12(4), 5-10. https://europub.co.uk/articles/-A-522525