Validation of Hemochroma PLUS: A Point of Care Testing Device for Haemoglobin Estimation

Journal Title: Journal of Clinical and Diagnostic Research - Year 2019, Vol 13, Issue 5

Abstract

ABSTRACT Introduction: Anaemia is a major public health challenge in India. Advent of Point of Care (POC) testing devices for haemoglobin estimation may be useful in the field settings, albeit after proper validation of these devices. Aim: To assess the validity of Hemochroma PLUS in haemoglobin estimation by comparing it with an automated laboratory analyser in rural Indian context. Materials and Methods: A cross-sectional study was conducted at a tertiary care rural teaching hospital during November-December 2017. Adults visiting the out-patient laboratory collection unit for a haemogram test were included in the study. Venous blood samples were collected by trained phlebotomists and tested using autoanalyser in the laboratory. Capillary blood samples were collected by a trained study coordinator using a lancet (Ez-life) and tested on a Hemochroma PLUS device. Haemoglobin measurement by both methods was compared using Bland-Altman analysis. The analysis was performed using STATA (14.2). Results: A total of 100 (53 females and 47 males) adults participated in the study. Mean (SD) age of the participants was 44.62 (16.92) years. Bland-Altman analysis revealed that the mean difference (95% confidence limits) was 0.3 (-2.0, 2.7) gm/dL. Hemochroma PLUS was found to have sensitivity of 84.90% and specificity of 82.98% in identifying anaemia. Conclusion: Hemochroma PLUS showed good agreement with the cell counter in current study. Further studies may be undertaken in different study populations and study settings. Hemochroma PLUS, has a potential as a POC tool for community-based anaemia screening and evaluation of anaemia control programmes.

Authors and Affiliations

Shyamsundar J Raithatha, Mustafa F Ranapurwala, Simmy Lahori, Chandrashekhar N Bopche, Ajay Gajanan Phatak

Keywords

Related Articles

Serum Zinc Levels in Thai Children with Acute Diarrhoea

ABSTRACT Introduction: Diarrhoea remains a leading cause of morbidity and mortality in children in developing countries. Zinc has been recommended by the WHO for the prophylaxis and treatment of acute diarrhoea. However,...

Hemangiopericytoma: A Rare Mass Arising in the Kidney

ABSTRACT Hemangiopericytoma, a tumour arising from pericytes is an unusually rare kidney neoplasm. It is difficult to establish a preoperative diagnosis. Here, we present a case report of a 47-year-old gentleman who pres...

Filarial Pleural Effusion with Lymphoma: A Rare Association

ABSTRACT Filariasis is endemic in developing countries with highest prevalence in India. Microfilarias are infrequently reported during cytological examination of body cavity fluids. Its association with neoplasm is stil...

A Misdiagnosed Laryngeal Web: Treated with an Innovative Self-Made Keel

Laryngeal web is a rare congenital anomaly. Late presentation at the age of 10 years as in our case, with misleading diagnosis of asthma in a Cohen’s type III glottic web is even rarer. In cases of congenital laryngeal w...

Dental Students’ Perception of Dentine Hypersensitivity and Awareness about its Management

ABSTRACT Introduction: Dentine hypersensitivity is a common painful oral condition that can interfere with drinking, eating, tooth brushing and even breathing. Successful management depends on the identification and elim...

Download PDF file
  • EP ID EP584550
  • DOI 10.7860/JCDR/2019/39995.12824
  • Views 161
  • Downloads 0

How To Cite

Shyamsundar J Raithatha, Mustafa F Ranapurwala, Simmy Lahori, Chandrashekhar N Bopche, Ajay Gajanan Phatak (2019). Validation of Hemochroma PLUS: A Point of Care Testing Device for Haemoglobin Estimation. Journal of Clinical and Diagnostic Research, 13(5), 7-9. https://europub.co.uk/articles/-A-584550