Acceptance Tests and Quality Control of the Positron Emission Tomography (PET) Systems

Journal Title: Nuclear Medicine Seminars - Year 2020, Vol 6, Issue 2

Abstract

Quality control (QC) applications are important for the reliability and accuracy of the results obtained from positron emission tomography (PET) imaging systems in clinical trials. With these applications, exposure of patients and employees to more radiation will be prevented, and it will be possible to detect the systems that lost their performance compared to their installation time. QC applications are grouped as acceptance and reference tests, routine periodic QC tests. PET performance tests standardized by National Electrical Manufacturers Association (NEMA) should be used as acceptance and reference tests. In addition, the performance of the device should be monitored by following these tests at six-month intervals. PET/computerized tomography (CT) manufacturers often recommend routine QC procedures for their equipment and require users to follow these recommendations to a minimum. If standards are not defined for routine QC procedures, it is recommended to comply with the general minimum standards for routine QC that all PET/CT centers must perform. Routine QC protocols, the image quality of the PET system; CT image quality and CT dose of the patient; the accuracy of CT-based attenuation corrections; the accuracy of the CT and PET co-registration should ensure that problems that can interact any of the performance characteristics are identified. Since performing PET/CT scans by equipment that does not comply with their specifications may compromise image quality, interpretation of images and reliability of results, daily QC parameters must be evaluated and necessary corrective procedures must be taken prior to scanning. The tests are carried out using the protocol provided by the manufacturer.

Authors and Affiliations

Tuğba Hacıosmanoğlu, Mustafa Demir, Türkay Toklu, F. Suna Kıraç, Mehmet İnce, Yasemin Parlak, Bilal Kovan, Semra Dönmez, Aslı Ayan, İnci Alıç Özaslan, Meral Hiçürkmez, Nami Yeyin, Leyla Poyraz ve TNTD Kalite kontrol ve Radyasyon Güvenliği çalışma grubu üyeleri

Keywords

Related Articles

Renal Cortical Scintigraphy Procedure Guideline

This guideline is the updated version of TNTD renal cortical scintigraphy guideline published in 2001. It includes the current recommended patient doses and patient dosimetry according to international guidelines.

Differentiated Thyroid Carcinoma in Pediatric Age Group/Differences in Diagnosis, Therapy and Follow-up

Differentiated thyroid cancer in childhood is rare. Although its clinical course is quite aggressive with higher rates of lymph node and pulmonary metastases as compared to adults, the final outcome tends to be favorable...

Radiation Safety Guide: General Definitions and Radiation Protection Rules in Nuclear Medicine Applications

Ionizing radiation which is widely used in the field of medicine and industry, and in scientific studies leaves some or all of its energy to the tissue as it passes through living tissue. It causes known possible harmful...

Non-ultrasonographic and Non-radioiodine Imaging Techniques in Thyroid Cancer

Ultrasonography is the most commonly used and most effective imaging modality in thyroid cancer imaging. It is the most successful method for evaluating primary tumor and local disease. I-131 scintigraphy is also a usefu...

Myocardial Perfusion SPECT Procedure Guideline

After Turkish Society of Nuclear Medicine “Cardiology Group” had first published “Nuclear Cardiology Practice Guidelines” in 2001, in line particular with the technological developments, significant changes both in the h...

Download PDF file
  • EP ID EP688741
  • DOI 10.4274/nts.galenos.2020.0008
  • Views 157
  • Downloads 0

How To Cite

Tuğba Hacıosmanoğlu, Mustafa Demir, Türkay Toklu, F. Suna Kıraç, Mehmet İnce, Yasemin Parlak, Bilal Kovan, Semra Dönmez, Aslı Ayan, İnci Alıç Özaslan, Meral Hiçürkmez, Nami Yeyin, Leyla Poyraz ve TNTD Kalite kontrol ve Radyasyon Güvenliği çalışma grubu üyeleri (2020). Acceptance Tests and Quality Control of the Positron Emission Tomography (PET) Systems. Nuclear Medicine Seminars, 6(2), -. https://europub.co.uk/articles/-A-688741