An Analytical- empirical Calculation of Linear Attenuation Coefficient of Megavoltage Photon Beams
Journal Title: Journal of Biomedical Physics and Engineering - Year 2017, Vol 7, Issue 3
Abstract
Background: In this study, a method for linear attenuation coefficient calculation was introduced. Methods: Linear attenuation coefficient was calculated with a new method that base on the physics of interaction of photon with matter, mathematical calculation and x-ray spectrum consideration. The calculation was done for Cerrobend as a common radiotherapy modifier and Mercury. Results: The values of calculated linear attenuation coefficient with this new method are in acceptable range. Also, the linear attenuation coefficient decreases slightly as the thickness of attenuating filter (Cerrobend or mercury) increased, so the procedure of linear attenuation coefficient variation is in agreement with other documents. The results showed that the attenuation ability of mercury was about 1.44 times more than Cerrobend. Conclusions: The method that was introduced in this study for linear attenuation coefficient calculation is general enough to treat beam modifiers with any shape or material by using the same formalism; however, calculating was made only for mercury and Cerrobend attenuator. On the other hand, it seems that this method is suitable for high energy shields or protector designing.
Authors and Affiliations
F Seif, M J Tahmasebi-Birgani, M R Bayatiani
Synergistic Effect of Radiofrequency Electromagnetic Fields of Dental Light Cure Devices and Mobile Phones Accelerates the Microleakage of Amalgam Restorations: An in vitro Study
Background: Previous studies have shown that exposure to electromagnetic fields produced by magnetic resonance imaging or mobile phones can lead to increased microleakage of dental amalgam. Objective: The aim of the pres...
Calculating Weighting Factors for Mixing Megavoltage Photon Beams to Achieve Desirable Dose Distribution in Radiotherapy
Background: In radiotherapy, low-energy photon beams are better adapted to the treated volume, and the use of high-energy beams can reduce hot spots in the radiation therapy. Therefore, mixing low and high energies with...
Radiation-induced Non-targeted Effect and Carcinogenesis; Implications in Clinical Radiotherapy
Bystander or non-targeted effect is known to be an interesting phenomenon in radiobiology. The genetic consequences of bystander effect on non-irradiated cells have shown that this phenomenon can be considered as one of...
Heart Rate Variability Classification using Support Vector Machine and Genetic Algorithm
Background: Electrocardiogram (ECG) is defined as an electrical signal, which represents cardiac activity. Heart rate variability (HRV) as the variation of interval between two consecutive heartbeats represents the balan...
An Analytical- empirical Calculation of Linear Attenuation Coefficient of Megavoltage Photon Beams
Background: In this study, a method for linear attenuation coefficient calculation was introduced. Methods: Linear attenuation coefficient was calculated with a new method that base on the physics of interaction of photo...