Application of Sigma Metrics and Performance Comparison Between Two Biochemistry Analyser and A Blood Gas Analyser for the Determination of Electrolytes

Journal Title: Journal of Clinical and Diagnostic Research - Year 2017, Vol 11, Issue 2

Abstract

Introduction: Electrolytes have a narrow range of biological variation and small changes are clinically significant. It is important to select the best method for clinical decision making and patient monitoring in the emergency room. The sigma metrics model provides an objective method to evaluate the performance of a method. Aim: To calculate sigma metrics for electrolytes measured with one arterial blood gas analyser including two auto-analysers that use different technologies. To identify the best approach for electrolyte monitoring in an emergency setting and the context of routine emergency room workflow. Materials and Methods: The Coefficient of Variation (CV) was determined from Internal Quality Control (IQC). Data was measured from July 2015 to January 2016 for all three analysers. The records of KBUD external quality data (Association of Clinical Biochemists, Istanbul, Turkey) for both Mindray BS-2000M analyser (Mindray, Shenzhen, China) and Architect C16000 (Abbott Diagnostics, Abbott Park, IL) and MLE clinical laboratory evaluation program (Washington, DC, USA) for Radiometer ABL 700 (Radiometer Trading, Copenhagen, Denmark) during the study period were used to determine the bias. Results: The calculated average sigma values for sodium (-1.1), potassium (3.3), and chloride (0.06) were with the Radiometer ABL700. All calculated sigma values were better than the autoanalysers. Conclusion: The sigma values obtained from all analysers suggest that running more controls and increasing the calibration frequency for electrolytes is necessary for quality assurance.

Authors and Affiliations

Yasemin Ustundag- Budak, Kagan Huysal

Keywords

Related Articles

Evaluation of Use of Adipose Cells in Various Procedures of Plastic Surgery and its Volumetric Quantification using High Resolution Ultrasound

Introduction: Autologous fat transfer is based on one of the principles of plastic surgery, replace like with like. It is used in wide variety of aesthetic procedures such as lip, facial augmentation and reconstructive p...

Inguinal Hernia as an Initial Presentation of Pseudomyxoma Peritonei

Pseudomyxoma Peritonei (PMP) is characterized by presence of abundant extracellular mucin in the peritoneum. Preoperative diagnosis of pseudomyxoma peritonei is difficult due to its varied presentation. We report a case...

A Rare Combination of Genitourinary Vascular Anomaly

ABSTRACT Vascular anatomy of the genitourinary system is essential to understand the complications related to surgeries, as it may influence the blood flow to the kidney and to the gonads. In the present case report, ar...

Effect of Subgingival Irrigation with Natural Products as an Adjunct to Scaling and Root Planing in the Treatment of Chronic Periodontitis- A Systematic Review

ABSTRACT Introduction: Various chemotherapeutic agents such as ‘Chlorhexidine’ (CHX), ‘Metronidazole’, ‘Povidone-iodine’ are used as a subgingival irrigant for management of periodontitis, but they have their own drawbac...

Pulpal Response to Ferric Sulfate and Diode Laser When Used as Pulpotomy Agent: An In vivo Study

Introduction: Ferric sulfate and Laser has been used for conventional pulpotomy therapy. Clinical and radiographic evaluation of pulpotomies using these have been done in numerous studies but there exists few studies ana...

Download PDF file
  • EP ID EP331844
  • DOI 10.7860/JCDR/2017/23486.92
  • Views 105
  • Downloads 0

How To Cite

Yasemin Ustundag- Budak, Kagan Huysal (2017). Application of Sigma Metrics and Performance Comparison Between Two Biochemistry Analyser and A Blood Gas Analyser for the Determination of Electrolytes. Journal of Clinical and Diagnostic Research, 11(2), 6-9. https://europub.co.uk/articles/-A-331844