Comparison of immersion ultrasound and low coherence reflectometry for ocular biometry in cataract patients

Journal Title: International Journal of Ophthalmology - Year 2018, Vol 11, Issue 6

Abstract

"AIM: To compare the results of axial length (AL) biometry in cataract eyes by three methods: immersion B-ultrasound (IB) biometry, immersion A-ultrasound (IA) biometry and optical low coherence reflectometry. METHODS: In this prospective observational study of eyes with cataract AL measurements were performed using immersion ultrasound and optical low coherence reflectometry device. The results were evaluated using Bland-Altman analyses. The differences between both methods were assessed using the paired t-test, and its correlation was evaluated by Pearson coefficient. RESULTS: Eighty eyes of 80 patients (39 men and 41 women) for cataract surgery were included in the study. The values of AL could be got from all 80 eyes by IB and IA, the difference of AL measurements between IA and IB was of no statistical significance (P=0.97); the mean difference in AL measurements was -0.031 mm (P=0.26; 95%CI, -0.09 to 0.02); linear regression showed an excellent correlation (r=0.98, P<0.0001). Forty-five of eighty eyes with results of AL measurements, which can be obtained by three methods; the difference of AL measurements was of no statistical significance (IA vs IB, P=0.18; IA vs Lenstar, P=0.51; IB vs Lenstar, P=0.07); linear regression showed an excellent correlation (IA vs IB, r=0.99; IA vs Lenstar, r=0.96; IB vs Lenstar, r=0.96); Bland-Altman analysis also showed good agreement between the two methods [IA vs IB, 95% limits of agreement (LoA), -0.36 to 0.28 mm; IA vs Lenstar, 95% LoA, -0.65 to 0.69 mm; IB vs Lenstar, 95% LoA, -0.55 to 0.68 mm]. CONCLUSION: Measurements with the optical low coherence reflectometry correlated well with IB and IA. In the eyes with serious refractive medium opacity, the measurements of AL could not be achieved or existed deviations when using optical low coherence reflectometry device. Under such circumstances, we should choose IA or IB as the optimization method to obtain measurements, in order to get much more accurate results."

Authors and Affiliations

Xiao-Yong Yuan

Keywords

Related Articles

Effects of exogenous recombinant human bone morphogenic protein-7 on the corneal epithelial mesenchymal transition and fibrosis

"AIM: To evaluate the effect of exogenous recombinant human bone morphogenic protein-7 (rhBMP-7) on transforming growth factor-β (TGF-β)-induced epithelial mesenchymal cell transition (EMT) and assessed its antifibrotic...

Efficacy and safety of ranibizumab for wet age-related macular degeneration in Chinese patients

"AIM: To evaluate the clinical efficacy and safety of ranibizumab for wet age-related macular degeneration (wAMD) in Chinese patients and to determine the mean number of injections administered over one year of follow-up...

miR-211 promotes lens epithelial cells apoptosis by targeting silent mating-type information regulation 2 homolog 1 in age-related cataracts

"AIM: To detect the expression of miR-211 in age-related cataract tissue, explore the effects of miR-211 on lens epithelial cell proliferation and apoptosis, and identify its target gene. METHODS: This study used real-ti...

Modified ulcer debridement in the treatment of the superficial fungal infection of the cornea

"AIM: To evaluate the efficacy of modified corneal ulcer debridement in superficial fungal keratitis unresponsive to medications. METHODS: A total of 209 patients (209 eyes) with fungal keratitis, involving no more than...

Download PDF file
  • EP ID EP484169
  • DOI 10.18240/ijo.2018.06.11
  • Views 47
  • Downloads 0

How To Cite

Xiao-Yong Yuan (2018). Comparison of immersion ultrasound and low coherence reflectometry for ocular biometry in cataract patients. International Journal of Ophthalmology, 11(6), -. https://europub.co.uk/articles/-A-484169