Evaluation of Mechanical Performance of Orthodontic Mini-Implants With Distinct Designs

Journal Title: UNKNOWN - Year 2017, Vol 12, Issue 1

Abstract

Background Orthodontic mini-implants represent one of the most used anchorage systems in daily clinical practice and frequently, these anchorage devices have to be placed between the roots of adjacent teeth reflecting the need to limit its dimensions. Among other factors, it has been suggested that mini-implant’s geometry have a great impact on primary stability. Objectives The aim of this study was to evaluate the primary stability (PS) of commercially available orthodontic mini-implants (MI) of different geometrical designs. Methods Twenty self-drilling MI (1.6 mm × 8 mm) were allocated to two groups (n = 10) according to their geometric design: INP-1 and INP-2. Twenty sections (8 mm ø × 10 mm long) were taken from bovine pelve, immersed in saline solution and stored by freezing (-20°C). Geometric characteristics concerning shape, diameter, length, pitch, thread details (number, angle and depth), thread shape factor (TSF) and conicity were evaluated by using Scanning Electron Microscope (SEM) imaging and all the measurements were carried out with Image-Pro® Insight Software. PS was evaluated by insertion torque (IT) and Periotest (PTV). Intergroup comparisons were made by Student t test. The level of significance was 5%. Results Increased IT values were observed for INP-1 (20.08 ± 3.28 N.cm) (P < 0.05), the group that presented the greatest conicity (33%) and highest thread depth (0.19 mm). Conclusions The wide geometric variability of MI indicates that MI design parameters affect its PS and it could be selected in order to improve its mechanical performance.

Authors and Affiliations

Amanda Carneiro da Cunha, Mariana Marquezan, Lincoln Issamu Nojima, Eduardo Franzotti Sant’Anna

Keywords

Related Articles

Zirconia Surface Treatment by Different Output Powers of Er: YAG Laser and Sandblasting: SEM Evaluation

Background: Application of zirconia in dentistry has increased due to its good properties. Since zirconia is not possible to be etched, evaluation of the other methods of surface treatment is important. Objectives: The...

Comparison of Maxillary Sinus Sizes in Patient with Maxillary Excess and Maxillary Deficiency

Objectives The aim of this study was to compare the maxillary sinus sizes in patients with maxillary excess...

Is Early Treatment With Functional Appliances Worth the Effort? A Discussion of the Pros and Cons of Early Interceptive Treatment

Early treatment of Class II, Div. 1 malocclusion is a much debated approach and some evidenced based studi...

Cephalometric Comparison of Position of the Hyoid Bone in Class I and Class II Patients

Background According to the close proximity of hyoid bone with dentofacial structures and its muscular atta...

Determination of Specific Anatomic Locations by SIFT Algorithm for Locating Cephalometric Points with the Use of an Automatic Technique

Background: Lateral cephalometry is very important for the evaluation of growth, diagnosis, treatment planning and therapeutic evaluations. Considering the advantages of lateral cephalometric analyses with the automatic...

Download PDF file
  • EP ID EP196483
  • DOI 10.17795/ijo-6527
  • Views 96
  • Downloads 0

How To Cite

Amanda Carneiro da Cunha, Mariana Marquezan, Lincoln Issamu Nojima, Eduardo Franzotti Sant’Anna (2017). Evaluation of Mechanical Performance of Orthodontic Mini-Implants With Distinct Designs. UNKNOWN, 12(1), -. https://europub.co.uk/articles/-A-196483