Geometrical Precision and Surface Topography of mSLA-Produced Surgical Guides for the Knee Joint

Journal Title: Journal of Engineering Management and Systems Engineering - Year 2023, Vol 2, Issue 3

Abstract

In this study, the precision of anatomical models and surgical guides pertaining to the knee joint, fabricated using the mSLA technique, was critically examined. The ITK-SNAP program was employed for the segmentation and reconstruction of knee joint anatomical structures, while surgical guide modelling was executed using the Siemens NX program. Subsequent fabrication of the models was accomplished with the Anycubic Photon Mono 4K MSLA 3D printer. An MCA II articulated arm equipped with a laser head, in conjunction with a TalyScan 150 profilometer, was utilized to gauge both the geometrical fidelity and surface roughness of the resulting models. Results indicated that the geometrical precision of these models remained within a tolerance of +/-0.3 mm. With regard to surface roughness, the Sa parameter was observed to lie between 2 and 2.5 µm.

Authors and Affiliations

Paweł Turek, Jakub Jakubiec

Keywords

Related Articles

The Impact of the COVID-19 Pandemic on Software Business Enterprises in Pakistan: Analysis and Implications

The COVID-19 pandemic emerged over three years ago as a public health crisis, swiftly evolving into a worldwide economic crisis with far-reaching implications for global business enterprises and industries. The unprecede...

Analysis and Regulation of Mechatronic Systems in Advanced Mobile Machines

In modern mobile machines, mechatronic systems have been integrated, enabling: a) the automation and robotization of machine tasks, b) the regulation of drive system parameters, and c) the transfer and processing of sign...

Applications of Machine Learning in Aircraft Maintenance

Aircraft maintenance is an expansive multidisciplinary field which entails robust design and optimization of extensive maintenance operations and procedures; encompassing the fault identification, detection and rectifica...

Enhanced Acoustic Attenuation Performance of a Novel Absorptive Muffler: A Helmholtz Equation-Based Simulation Study

This investigation delves into the noise attenuation capabilities of an innovatively designed muffler, which integrates additional piping and perforation to augment sound reflection. The enhanced muffler's design was rig...

Efficacy and Space Optimization in Industrial Warehouses: An Evaluation of Paternoster Continuous Vertical Conveyors

In the field of industrial buildings, notably within warehouse settings, the optimization of floor space emerges as a paramount concern. The deployment of equipment facilitating continuous transport is mandated to not on...

Download PDF file
  • EP ID EP731796
  • DOI 10.56578/jemse020302
  • Views 43
  • Downloads 0

How To Cite

Paweł Turek, Jakub Jakubiec (2023). Geometrical Precision and Surface Topography of mSLA-Produced Surgical Guides for the Knee Joint. Journal of Engineering Management and Systems Engineering, 2(3), -. https://europub.co.uk/articles/-A-731796