Neutron dose evaluation of Elekta Linac at two energies (10 & 18 MV) by MCNP code and comparison with experimental measurements
Journal Title: JOURNAL OF ADVANCES IN PHYSICS - Year 0, Vol 6, Issue 1
Abstract
Medical linear accelerators, besides the clinically high energy electron and photon beams, produce other secondary particles such as neutrons which escalate the delivered dose. In this study the neutron dose at 10 and 18MV Elekta linac was obtained by using TLD600 and TLD700 as well as Monte Carlo simulation. For neutron dose assessment in 2020 cm2 field, TLDs were calibrated at first. Gamma calibration was performed with 10 and 18 MV linac and neutron calibration was done with 241Am-Be neutron source. For simulation, MCNPX code was used then calculated neutron dose equivalent was compared with measurement data. Neutron dose equivalent at 18 MV was measured by using TLDs on the phantom surface and depths of 1, 2, 3.3, 4, 5 and 6 cm. Neutron dose at depths of less than 3.3cm was zero and maximized at the depth of 4 cm (44.39 mSvGy-1), whereas calculation resulted in the maximum of 2.32 mSvGy-1 at the same depth. Neutron dose at 10 MV was measured by using TLDs on the phantom surface and depths of 1, 2, 2.5, 3.3, 4 and 5 cm. No photoneutron dose was observed at depths of less than 3.3cm and the maximum was at 4cm equal to 5.44mSvGy-1, however, the calculated data showed the maximum of 0.077mSvGy-1 at the same depth. The comparison between measured photo neutron dose and calculated data along the beam axis in different depths, shows that the measurement data were much more than the calculated data, so it seems that TLD600 and TLD700 pairs are not suitable dosimeters for neutron dosimetry in linac central axis due to high photon flux, whereas MCNPX Monte Carlo techniques still remain a valuable tool for photonuclear dose studies.
Authors and Affiliations
Negin Shagholi, Hassan Ali, Mahdi Sadeghi, Arjang Shahvar, Hoda Darestani, Banaee Nooshin, Kheirolah Mohammadi
Power of Magnetic
The theme here is about magnetism and what is produced by it, looking at the sun, the earth, as well as the atom, and in general. It also shows magnetic for its nature of - with or without pressure to limit the velocity...
Effect of Annealing Temperature on The Refrective Index and Dielectric Properties of TiO2/MnO2 CORE/SHELL Thin Films
Thin film of the form TiO2/MnO2 was deposited using the chemical bath method. The deposited thin films were annealed at temperatures of in order to investigate the effect of annealing temperature on the refractive index...
Effect of Magnetic Field From Mobile Phone on Brain
Human exposure to electromagnetic field (EMF) comes from many different sources and occurs in various situations in everyday life. Man-made static fields are mainly found in occupational settings, such as close to magnet...
Proof of Renormalizability of Scalar Field Theories Using the Epstein-Glaser Scheme and Techniques of Microlocal Analysis
The renormalizability of QFT's is a vastly studied issue, and particularly the results concerning a scalar eld theory are well-known through the traditional renormalization approach in the literature. However, in this pa...
Path to Lasting Peace
Maybe many a time you looked at clock and realized passing time as normal or according to Einstein or Heisenberg theory have a deeper view about time and place acting. I was thinking about this great acting as you and al...