Optimization design and performance study of magnesium alloy vascular clamp

Journal Title: Progress in Medical Devices - Year 2024, Vol 2, Issue 3

Abstract

Purpose: This study investigates the effects of varying inner diameter of a vascular clamp made from an Mg–Nd–Zn–Zr alloy on its functional performance. The primary objectives are to optimize the clamp’s structure, assess its performance across different inner diameters, and ultimately determine the optimal configuration. Methods: We developed a V-shaped vascular clamp equipped with a locking mechanism and transverse teeth. The study involved comparing vascular clamps with various inner diameters (0.35 mm, 0.4 mm, 0.5 mm and 0.6 mm, denoted as R0.35, R0.4, R0.5 and R0.6, respectively), achieved by modifying the clamp design. Finite element analysis simulated the closure process of these clamps, both with and without blood vessels, to analyze stress and strain distribution. Subsequently, we manufactured a clamp with the optimized design and conducted performance evaluations, including a closing strength test and an in vitro immersion test. Results: Among the tested vascular clamps, the R0.5 clamp demonstrated the lowest strain (0.50798) and minimal stress on blood vessels (0.7629 MPa). Notably, the R0.5 clamp remained intact during clamping fracture experiments and demonstrated a maximum closing force of 334.98±15.4 mmHg. Regarding corrosion resistance, the clamped position showed a higher corrosion rate (0.179±0.00551 mg.cm-2.day-1) compared to the open clamp (0.161±0.00306 mg.cm-2.day-1). Conclusion: The R0.5 clamp demonstrated superior performance in finite element analysis, showing effective vascular closure, strong clamping force, and uniform corrosion behavior. Overall, these results highlight its potential as an effective tool for vascular closure.

Authors and Affiliations

Weiwei Fan, Lin Mao, Bojun Liu, Chengli Song

Keywords

Related Articles

Research progress on biliary stents

Bile duct stenosis is a common condition in gastroenterology and hepatobiliary surgery and can be divided into benign stenosis and malignant stenosis according to different etiologies. The implantation of a gall stent in...

Progress on Microfluidic Blood Cell Counting Techniques

Accurately and efficiently detecting the quantity of blood cells is crucial in routine blood examination, as abnormal high or low numbers of blood cells are associated with the occurrence of various disorders. Due to inh...

Integrating Traditional Chinese Medicine massage therapy with machine learning: A new trend in future healthcare

The growing demand for healthcare has brought Traditional Chinese Medicine (TCM) massage therapy into the spotlight in academic circles. Numerous studies have underscored the effectiveness of TCM massage in health promot...

Analysis and modeling of forced-damped vibrations and their applications in medicine

Forced-damped vibrations are pivotal in various medical applications, significantly contributing to the examination of tissue mechanical properties, development of medical devices, and understanding of biological systems...

Preparation and application of electro-conductive hydrogels in biomedical engineering

Electro-conductive hydrogel is a new composite hydrogel with high electrical conductivity, extraordinary mechanical properties, and controllability. It can be equipped with specific materials to obtain specific propertie...

Download PDF file
  • EP ID EP750327
  • DOI 10.61189/883654uegazz
  • Views 29
  • Downloads 0

How To Cite

Weiwei Fan, Lin Mao, Bojun Liu, Chengli Song (2024). Optimization design and performance study of magnesium alloy vascular clamp. Progress in Medical Devices, 2(3), -. https://europub.co.uk/articles/-A-750327