Optimization design and performance study of magnesium alloy vascular clamp

Journal Title: Progress in Medical Devices - Year 2024, Vol 2, Issue 3

Abstract

Purpose: This study investigates the effects of varying inner diameter of a vascular clamp made from an Mg–Nd–Zn–Zr alloy on its functional performance. The primary objectives are to optimize the clamp’s structure, assess its performance across different inner diameters, and ultimately determine the optimal configuration. Methods: We developed a V-shaped vascular clamp equipped with a locking mechanism and transverse teeth. The study involved comparing vascular clamps with various inner diameters (0.35 mm, 0.4 mm, 0.5 mm and 0.6 mm, denoted as R0.35, R0.4, R0.5 and R0.6, respectively), achieved by modifying the clamp design. Finite element analysis simulated the closure process of these clamps, both with and without blood vessels, to analyze stress and strain distribution. Subsequently, we manufactured a clamp with the optimized design and conducted performance evaluations, including a closing strength test and an in vitro immersion test. Results: Among the tested vascular clamps, the R0.5 clamp demonstrated the lowest strain (0.50798) and minimal stress on blood vessels (0.7629 MPa). Notably, the R0.5 clamp remained intact during clamping fracture experiments and demonstrated a maximum closing force of 334.98±15.4 mmHg. Regarding corrosion resistance, the clamped position showed a higher corrosion rate (0.179±0.00551 mg.cm-2.day-1) compared to the open clamp (0.161±0.00306 mg.cm-2.day-1). Conclusion: The R0.5 clamp demonstrated superior performance in finite element analysis, showing effective vascular closure, strong clamping force, and uniform corrosion behavior. Overall, these results highlight its potential as an effective tool for vascular closure.

Authors and Affiliations

Weiwei Fan, Lin Mao, Bojun Liu, Chengli Song

Keywords

Related Articles

Control technologies of lower limb rehabilitation exoskeleton robots based on surface electromyography: A review

The aging population is accompanied by a decline in human body function, leading to an increasing number of people with lower limb dysfunction, which has become a global public health challenge today. The lower limb reha...

Application and progress of functionalized magnetic bead-based biosensors for protein detection

In the field of bioanalysis, the integration of magnetic beads and biosensors provides a protein detection platform with high separation efficiency and sensitivity. The superparamagnetism of magnetic beads, combined with...

Progress of near-infrared spectroscopy in cerebral blood oxygenation detection: A mini review

In contrast to conventional oximeters, near-infrared spectroscopy-based brain tissue oximetry monitoring devices are capable of non-invasive, continuous, and real-time quantitative monitoring of cerebral oximetry paramet...

Optimal design and experimental study of Mg alloy electrodes for tissue welding

Objective: To investigate electrode weight reduction through structural modifications, focusing on variations in the thickness of circular electrodes to meet mechanical requirements. Additionally, optimal welding paramet...

Design and simulation of lower limb exoskeleton based on online gait generation algorithm

Objective: This study presents the design of an innovative lower limb exoskeleton featuring dynamic self-balancing capabilities. It employs Zero Moment Point and Model Predictive Control online gait algorithms to plan st...

Download PDF file
  • EP ID EP750327
  • DOI 10.61189/883654uegazz
  • Views 38
  • Downloads 0

How To Cite

Weiwei Fan, Lin Mao, Bojun Liu, Chengli Song (2024). Optimization design and performance study of magnesium alloy vascular clamp. Progress in Medical Devices, 2(3), -. https://europub.co.uk/articles/-A-750327