Performance verification and sigma metrics of creatinine assays

Journal Title: International Journal of Clinical Biochemistry and Research - Year 2018, Vol 5, Issue 1

Abstract

Introduction: Creatinine enzymatic method is more accurate but its own higher cost is the main reason why the use of Jaffe assays is still in practice. The verification step is performed because assay procedure confirms to the manufacturer’s claims in the end users’ setup. Desirable specification based on biological variation gives Allowable Total Error (ATE) and Sigma metrics provides summative evaluation of method performance. Objectives: Verification of manufacturer’s precision claims. Calculation of bias, Total Analytical Error (TAE) and sigma metrics. Materials and Methods: Modified Jaffe’s (No deproteinization) and enzymatic creatinine Erba XL liquid pack on Erba EM360 fully auto analyser were used. Intra assay and inter assay imprecision done on 20 replicates each. The estimates of bias uses data from precision experiment. Results & Discussion: The performance of jaffe’s and enzymatic serum creatinine is comparable to manufacturer's stated claimed value and are within desirable specification based on biological variation. The bias calculated in both assays is less than allowable bias. Total Analytical Error (TAE) is less than Allowable Total Error (ATE). Both the methods overall showed ATE ≥ bias + 4 SD. Within-run showed ATE ≥ bias + 5 SD complying Six Sigma concepts. Conclusion: Both Jaffe’s and enzymatic creatinine assays gives good performance and better sigma metrics. This makes enzymatic method a better choice due to inherent advantages such as increased accuracy and less interference if cost is neglected.

Authors and Affiliations

Santosh B. Jagtap, Sanjyoti S. Bandebuche, Mamata V. Hegde

Keywords

Related Articles

The synergistic role of CKMB mass and high sensitivity Troponin T in the diagnosis of Acute Coronary Syndrome

Aim: To determine the synergistic role of CKMB mass (CKMBM) and high sensitivity Troponin T (hs-cTnT) in the diagnosis of Acute Coronary Syndrome (ACS). Study Design: This was a cross-sectional study comprising of 140 in...

Correlation of anthropometric parameters with lipid profile in first year medical students

Introduction: Lifestyle related risk factors are prone to increase the burden of cardiovascular diseases. Thus, the study was carried out in the medical students who are more prone to develop lifestyle diseases because o...

A study comparing values of serum sodium estimated by colorimetric kit method with those obtained by direct and indirect ion selective electrode methods

Introduction Electrolyte abnormalities are known to cause significant morbidity and mortality when it is not detected Dyselectrolytemias are usually treatable Hyponatremia is a common manifestation encountered in the eld...

Thyroid dysfunction in patients with chronic kidney disease undergoing maintenance hemodialysis

Background: A thyroid dysfunction is a medical condition impairing the function of the thyroid. Chronic Kidney Disease (CKD) is a clinical syndrome that resulting from progressive loss of renal functions. A number of stu...

A comparative analysis of direct LDL-C assay and Friedewald’s formula in subjects of ischemic heart disease & stroke in a tertiary care centre

Introduction Among the various risk factors for the development of Ischemic heart diseases IHD as well as stroke is elevated low density lipoprotein cholesterol LDLC levels This makes accurate reporting of LDLC crucial i...

Download PDF file
  • EP ID EP472117
  • DOI 10.18231/2394-6377.2018.0024
  • Views 55
  • Downloads 0

How To Cite

Santosh B. Jagtap, Sanjyoti S. Bandebuche, Mamata V. Hegde (2018). Performance verification and sigma metrics of creatinine assays. International Journal of Clinical Biochemistry and Research, 5(1), 120-123. https://europub.co.uk/articles/-A-472117