Retrospective Data Analysis and Proposal of a Practical Acceptance Criterion for Inter-laboratory Cross-validation of Bioanalytical Methods Using Liquid Chromatography/Tandem Mass Spectrometry
Journal Title: The AAPS Journal - Year 2014, Vol 16, Issue 6
Abstract
The purpose of this study is to conduct a retrospective data analysis for inter-laboratory cross-validation studies to set a reasonable and practical acceptance criterion based on a number of cross-validation results. From the results of cross-validation studies for 16 compounds and their metabolites, analytical bias and variation were evaluated. The accuracy of cross-validation samples was compared with that of quality control (QC) samples with statistical comparison of the analytical variation. An acceptance criterion was derived with a confidential interval approach. As the results, while a larger bias was observed for the cross-validation samples, the bias was not fully caused by analytical variation or bias attributable to the analytical methods. The direction of the deviation between the cross-validation samples and QC samples was random and not concentration-dependent, suggesting that inter-laboratory variability such as preparation errors could be a source of bias. A derived acceptance criterion corresponds to one prescribed in the Guideline on bioanalytical method validation from the Ministry of Health, Labour and Welfare in Japan and is a little wider than one in the European Medical Agency. In conclusion, thorough retrospective data analysis revealed potential causes of larger analytical bias in inter-laboratory cross-validation studies. A derived acceptance criterion would be practical and reasonable for the inter-laboratory cross-validation study.
Authors and Affiliations
Tomoki Yoneyama, Takashi Kudo, Fumihiro Jinno, Eric R. Schmidt, Takahiro Kondo
Comparative Performance of Cell Life Span and Cell Transit Models for Describing Erythropoietic Drug Effects
The online version of this article (doi:10.1208/s12248-011-9302-9) contains supplementary material, which is available to authorized users.
Development, Validation, and Clinical Implementation of an Assay to Measure Total Antibody Response to Naglazyme® (Galsulfase)
Naglazyme® (galsulfase, rhASB) was developed as enzyme replacement therapy for mucopolysaccharidosis type VI. Naglazyme generated an IgG antibody response in most patients. To better characterize Naglazyme immuno...
Population pharmacokinetic studies in pediatrics: Issues in design and analysis
The current review addresses the following 3 frequently encountered challenges in the design and analysis of population pharmacokinetic studies in pediatrics: (1) body size adjustments during the development of pharmacos...
Effects of Device and Formulation on In Vitro Performance of Dry Powder Inhalers
The study examined the sensitivity of DPI in vitro performance to formulation and device changes. Rotahaler/Rotacaps was selected as the reference DPI drug product, and Aerolizer was selected as the test device. Since th...
2007 highlights of advances in the pharmaceutical sciences: An American Association of Pharmaceutical Scientists (AAPS) perspective
The American Association of Pharmaceutical Scientists (AAPS) covers the full range of areas of expertise associated with the resolution of concerns pertaining to drugs and drug products. This editorial highlights the ini...