Suitability of open-field autorefractors as pupillometers and instrument design effects

Journal Title: International Journal of Ophthalmology - Year 2017, Vol 10, Issue 4

Abstract

"AIM: To determine the agreement and repeatability of the pupil measurement obtained with VIP-200 (Neuroptics), PowerRef II (Plusoptix), WAM-5500 (Grand Seiko) and study the effects of instrument design on pupillometry. METHODS: Forty patients were measured twice in low, mid and high mesopic. Repeatability was analyzed with the within-subject standard deviation (Sw) and paired t-tests. Agreement was studied with Bland-Altman plots and repeated measures ANOVA. Instrument design analysis consisted on measuring pupil size with PowerRef II simulating monocular and binocular conditions as well as with proximity cues and without proximity cues. RESULTS: The mean difference (±standard deviation) between test-retest for low, mid and high mesopic conditions were, respectively: -0.09 (±0.16), -0.05 (±0.18) and -0.08 (±0.23) mm for Neuroptics, -0.05 (±0.17), -0.12 (±0.23) and -0.17 (±0.34) mm for WAM-5500, -0.04 (±0.27), -0.13 (±0.37) and -0.11 (±0.28) mm for PowerRef II. Regarding agreement with Neuroptics, the mean difference for low, mid and high mesopic conditions were, respectively: -0.48 (±0.35), -0.83 (±0.52) and -0.38 (±0.56) mm for WAM-5500, -0.28 (±0.56), -0.70 (±0.55) and -0.61 (±0.54) mm for PowerRef II. The mean difference of binocular minus monocular pupil measurements was: -0.83 (±0.87) mm; and with proximity cues minus without proximity cues was: -0.30 (±0.77) mm. CONCLUSION: All the instruments show similar repeat-ability. In all illumination conditions, agreement of Neuroptics with WAM-5500 and PowerRef II is not good enough, which can be partially induced due to their open field design."

Authors and Affiliations

Carles Otero

Keywords

Related Articles

Evaluation of systemic risk factors in different optical coherence tomographic patterns of diabetic macular edema

"AIM: To elucidate the relationship between systemic risk factors and different patterns of diabetic macular edema (DME) determined with optical coherence tomography (OCT). METHODS: In this cross-sectional study, DME was...

Infliximab relieves blood retinal barrier breakdown through the p38 MAPK pathway in a diabetic rat model

"AIM: To clarify the mechanism of infliximab treatment in diabetic macular edema (DME) and to provide a new alternative therapy for DME. METHODS: Rats were randomly divided into the control group, the model group and the...

Potential role of Müller cells in the pathogenesis of macropsia associated with epiretinal membrane: a hypothesis revisited

Pathophysiological explanations for metamorphopsia associated with retinal pathologies generally focus on photoreceptor organization disruption. However, the retinal microarchitecture is complicated, and we hypothesize t...

Comparison of anterior segment optical coherence tomography findings in acanthamoeba keratitis and herpetic epithelial keratitis

This study is to investigate the characteristic features of Acanthamoeba keratitis (AK) that differentiating it from herpetic epithelial keratitis (HEK) using anterior segment optical coherence tomography (AS-OCT). Medic...

Pediatric Nd:YAG laser capsulotomy in the operating room: review of 87 cases

"AIM: To report a large series of children having Nd:YAG laser capsulotomy in the operating room using the lateral decubitus position. METHODS: Medical records of children who underwent Nd:YAG laser capsulotomy in the op...

Download PDF file
  • EP ID EP605135
  • DOI 10.18240/ijo.2017.04.11
  • Views 48
  • Downloads 0

How To Cite

Carles Otero (2017). Suitability of open-field autorefractors as pupillometers and instrument design effects. International Journal of Ophthalmology, 10(4), -. https://europub.co.uk/articles/-A-605135