COMPARISON OF URISED 2 FULLY AUTOMATED URINE ANALYZER TO MANUAL URINE MICROSCOPY

Journal Title: PARIPEX-Indian Journal of Research - Year 2019, Vol 8, Issue 5

Abstract

Introduction: Urinalysis is one of the most commonly performed test in the clinical laboratory. Urine screening is achieved by either automated or manual microscopic analysis. The objective of this study was to compare the results of fully automated urine analyzer to manual urine microscopy. Material and Method : A total of 200 urine samples were analyzed by Urised 2 fully automated urine analyzer and manual method and the results were compared. Result : The sensitivity for RBC was 89.2% , specificity was 93.3% respectively. The sensitivity for WBC was 86% , specificity was 91% in the study conducted on 200 urine sample. Conclusion: The use of automated device in urinalysis is a good practice as it showed good compatibility to manual microscopy. However in order to avoid any error or uncertainity it requires careful evaluation and some images have to be analyzed by manual microscopic examination.

Authors and Affiliations

Dr Anjana Mittal, Dr. Shweta Sharma

Keywords

Related Articles

URINE CULTURE ISOLATES AND THEIR ANTIBIOTIC SENSITIVITY PATTERN IN A TERTIARY CARE HOSPITAL OF NORTH EAST INDIA

Urinary tract infection is one of the most common bacterial infections seen in clinical practice particularly in developing countries. The present study shall focus on microorganisms responsible for urinary tract infecti...

STUDY OF PREVALENCE OF ACCESSORY HEPATIC ARTERY IN 56 CADAVERS

INTRODUCTION- In modern surgical and transplantation procedures the recognition of anatomic vascular abnormalities of the hepatic arteries is of greater importance than ever. METHOD AND MATERIALS- This study was done on...

Healthy Foods to Control Cholesterol

Nowadays people realise that cholesterol is not good for human health. Increased blood cholesterol leads to blockage in the blood vessels of heart and brain. Too much of blockage can lead to heart attack or stroke. Human...

THE EXTENT TO WHICH RABINDRANATH TAGORE HAS SEEPED IN THE CULTURAL FABRIC OF THE Z GENERATION AS AN IMPETUS TO CULTURAL TOURISM IN SANTINIKETAN

As the century takes a turn, there is a new generation emerging whose characteristics are unique and much different from the Millennials and the Gen X. The ever increasing sector of Generation Z is now in focus, which is...

Iatrogenic hydatid cutaneous fistulization in a patient with coexisting pulmonary tuberculosis

We report a very rare case of Pleural hydatid cyst in a known case of pulmonary tuberculosis which presented as discharging wound of the chest wall resulting due to intercostal drain insertion misdiagnosing it as tubercu...

Download PDF file
  • EP ID EP586246
  • DOI -
  • Views 88
  • Downloads 0

How To Cite

Dr Anjana Mittal, Dr. Shweta Sharma (2019). COMPARISON OF URISED 2 FULLY AUTOMATED URINE ANALYZER TO MANUAL URINE MICROSCOPY. PARIPEX-Indian Journal of Research, 8(5), 12-13. https://europub.co.uk/articles/-A-586246