Laboratory Accuracy of Some Human Immunodeficiency Virus Screening Methods in a Nigerian Blood Bank: Is it time for Universal Adoption of Enzyme-linked Immuno-Sorbent Assay Methodologies as the Minimum Testing Paradigm?

Journal Title: International Blood Research & Reviews - Year 2015, Vol 3, Issue 2

Abstract

Aim: To compare the prevalence rates, relevant indices of laboratory accuracy and proportion of false negative test results for some WHO recommended methodologies used for HIV screening amongst blood donor sata hospital-based blood bank in Nigeria. Study Design: A cross-sectional. Place and Duration: Blood bank unit of Jos University Teaching Hospital (JUTH) and the Nigerian National Blood Transfusion Service (NBTS) North Central Zonal Office, Jos between May and August 2008. Methodology: Four hundred and forty blood donors (379 males and 61 females; aged 18-55 years) predominantly family replacement blood donors who met the minimum criteria to donate blood in Nigeria were included. Blood collection, serum processing, testing and interpretation of results were carried out using standard methods and manufacturers’ instruction. Serum was tested with a rapid test (Determine™ HIV- 1/2) and an EIA [Dia Pro HIV 1/2/0 ELISA] method. The samples were further tested with a 4th generation ELISA [GENSCREEN®PLUS HIV Ag- Ab ELISA]. Results: The prevalence of HIV in blood donors differed with the test method and assay as follows; Determine TM HIV 1/ 2 (3.6%), Dia Pro HIV 1/2/0 ELISA (5.5%) and GENSCREEN®PLUS HIV Ag-Ab ELISA (9.3) respectively. Determine TM HIV-1/ 2gave a sensitivity of 0.39 (95% CI 0.24-0.55), specificity 1.00,95% CI 0.99-1.00), false negative [FN] (61%), positive predictive value [PPV] 1.00 95% CI 0.79-1.00), and a negative predictive value [NPV] 0.94, 95% CI 0.91-0.96 when compared with GENSCREEN®PLUS HIV Ag-Ab ELISA method. P<0.001. Dia Pro HIV 1/2/0 ELISA gave a sensitivity of 0.54, 95% CI.37-0.69, specificity 0.995, 95% CI 0.99-1.00, FN(46.3%), PPV (0.9295% CI 0.73-0.99 and a NPV (0.95, 95% CI0.93-0.97) when compared with GENSCREEN®PLUS HIV Ag-Ab ELISA method. P<0.001. Determine TM HIV 1/2 had a sensitivity of 0.67 95% CI 0.45-0.84, specificity of 1.00; 95% CI 0.99-1.00, FN (33.3%), PPV (1.00 95% CI 0.79-1.00 and a NPV 0.98, 95% CI 0.96-0.99 when compared with Dia Pro HIV 1/2/0 ELISA method. P<0.001. Conclusion: The prevalence of HIV in blood donors is method dependent with GENSCREEN®PLUS HIV Ag-Ab ELISA higher than Dia Pro HIV 1/2/0 and Determine TM HIV 1/ 2. Dia Pro HIV 1/2/0 is more accurate and has fewer FN test results than Determine TM HIV 1/ 2. There is a need to discourage rapid testing as a major testing algorithm amongst hospital-based blood banks. Instead, ELISA methods should be adopted as the minimum testing paradigm. However, further testing with Nucleic Acid Amplification Testing (NAT) is recommended to validate reliability of this study.

Authors and Affiliations

Orkuma Joseph Aondowase, Gomerep Simji Samuel, Egesie Julie Ochaka, Mbaave Tsavyange Peter, Onoja Anthony Michael

Keywords

Related Articles

A Comparative Study on Hematological Inflammation Markers in Acne Vulgaris

Aim: The literature does not include sufficient data on the hematological inflammation markers in acne vulgaris. The present study aimed to investigate potential of neutrophil/lymphocyte ratio, platelet/lymphocyte ratio,...

Hematological Changes Following Oral Administration of Aqueous Root Bark Extract of Salacia lehmbachii in Albino Rats

Objective: This study investigated the effects of oral administration of aqueous root bark extract of Salacia lehmbachii on some haematological indices in albino rats. Methodology: Twenty-four male rats weighing 180-200...

A Preliminary Study of the Effect of Alchornea laxiflora on Haematological Parameters and Organ Weight in Albino Rats

Aim: The aim of this study is to investigate the effects of ethanol extract of Alchornea laxiflora on internal organs as well as haematological indices of rats. Methodology: The plant materials were air dried and grinded...

Artifactual Changes in Whole Blood and Plasma Glucose Levels of Diabetic and Non Diabetic Blood Samples Twenty Four Hours (24 h) Post Collection

Analysis of diabetic blood may be delayed for hours post collection owing to several factors. Artifactual changes may occur thereby confounding the results. This study investigated some artifactual changes in glucose lev...

Comparison Effect of Moringa oleifera Leaf Meal and Oxytetracycline on Haematology and Serum Biochemical Profile of Broiler Finishers

Aims: The study was carried out to assess the effects of Moringa oleifera leaf meal on the haematological parameters and serum biochemical profile of broiler finishers in comparison with oxytetracycline. Study Design:...

Download PDF file
  • EP ID EP242438
  • DOI 10.9734/IBRR/2015/17095
  • Views 102
  • Downloads 0

How To Cite

Orkuma Joseph Aondowase, Gomerep Simji Samuel, Egesie Julie Ochaka, Mbaave Tsavyange Peter, Onoja Anthony Michael (2015). Laboratory Accuracy of Some Human Immunodeficiency Virus Screening Methods in a Nigerian Blood Bank: Is it time for Universal Adoption of Enzyme-linked Immuno-Sorbent Assay Methodologies as the Minimum Testing Paradigm?. International Blood Research & Reviews, 3(2), 92-100. https://europub.co.uk/articles/-A-242438