The measurement of photoneutron dose around an 18 MV Varian linear accelerator by TLD600 and TLD700 dosimeters inside the polyethylene sphere of neutron probe LB 6411

Journal Title: JOURNAL OF ADVANCES IN PHYSICS - Year 0, Vol 6, Issue 1

Abstract

Bremsstrahlung X-rays from high-energy linear accelerators (Linacs) produce neutrons as a result ofphotonuclear reactions mainly with the different materials constituting the accelerator head. The neutronsproduced during high energy radiotherapy should be considered in terms of protection and ose escalation. Due to the very intense photon component in the Linac field causing pulse pile-up and dead-time effects indetectors, measurement of the corresponding neutron dose by active dosimeters is extremely troublesome. In this study, the neutron probe LB 6411, which active detector of 3He proportional counter tube was replaced withthe passive detectors of TLD600 and TLD700, has been used to perform neutron measurements at four pointsaround the 18 MV Varian 2100C Linac facility. The neutron dose equivalent at the distance of 1 and 2 m fromthe isocentre on the patient couch was obtained 2.2 and 0.75 mSv.Gy-1respectively. According to the results,the dose equivalent from emitted photoneutrons is not negligible and therefore treatment conditions should beoptimized. The results of this study emphasized that TLD600 and TLD700 dosimeters inside the polyethylenesphere of neutron probe LB 6411 is an appropriate choice for studying photoneutron production in the vicinity ofthe accelerator

Authors and Affiliations

Samiei F, Nedaie H A, Darestani H, Banaee N, Shagholi N, Mohammadi K, Shahvar A

Keywords

Related Articles

Simple analytical algorithm for constructing the envelope of the projectile trajectories in midair

A classic problem of the motion of a projectile thrown at an angle to the horizon is studied. The air drag force is taken into account asthe quadratic resistance law.An analytic approach is used for the investigation. Si...

Effect of quaternary addition on structure, electrical, mechanical and thermal properties of bismuth-tin-zinc rapidly solidified fusible alloy

Effect of quaternary addition, silver or indium, on structure, electrical, mechanical and thermal properties of bismuth-tin-zinc rapidly solidified fusible alloy have been investigated.  Adding silver or indium caus...

Thermodynamic Functions Molecular Polarizability of 2,6-dichloro-4-fluoro Phenol

The Laser-Raman and FT-IR spectra of 2,6-dichloro-4-fluoro phenol (2,6,4-DCFP) have been recorded. The thermodynamic functions, namely, the enthalpy, the heat capacity, the free energy and entropy of 2,6,4-DCFP&nbsp...

Dosimetric comparison and time consuming evaluations for neoadjuvant 5-fields versus 7-fields Intensity-Modulated Radiation Therapy for locally advanced rectal cancer.

Aims and background: Preoperative  multidrug - chemoradiotherapy schedule (PCR) is proven to be the standard approach in patients with locally advanced rectal cancer (LARC).  The main limiting factor of PC...

Evolution of unstable system.

Scenario of appearance and development of instability in problem of a flow around a solid sphere at rest is discussed. The scenario was created by solutions to the multimoment hydrodynamics equations, which were applied...

Download PDF file
  • EP ID EP653580
  • DOI 10.24297/jap.v6i1.1822
  • Views 54
  • Downloads 0

How To Cite

Samiei F, Nedaie H A, Darestani H, Banaee N, Shagholi N, Mohammadi K, Shahvar A (0). The measurement of photoneutron dose around an 18 MV Varian linear accelerator by TLD600 and TLD700 dosimeters inside the polyethylene sphere of neutron probe LB 6411. JOURNAL OF ADVANCES IN PHYSICS, 6(1), 1049-1057. https://europub.co.uk/articles/-A-653580