Enhancement of Toxic Substances Clearance from Blood Equvalent Solution and Human Whole Blood through High Flux Dialyzer by 1 MHz Ultrasound

Journal Title: Journal of Biomedical Physics and Engineering - Year 2017, Vol 7, Issue 2

Abstract

Background Hemodialysis is a process of removing waste and excess fluid from blood when kidneys cannot function efficiently. It often involves diverting blood to the filter of the dialysis machin to be cleared of toxic substances. Fouling of pores in dialysis membrane caused by adhesion of plasma protein and other toxins will reduce the efficacy of the filtre. Objectives In This study, the influence of pulsed ultrasound waves on diffusion and the prevention of fouling in the filter membrane were investigated. Methods Pulsed ultrasound waves with frequency of 1 MHz at an intensity of 1 W/cm2 was applied to the high flux (PES 130) filter. Blood and blood equivalent solutions were passed through the filter in separate experimental setups. The amount of Creatinine, Urea and Inulin cleared from both blood equvalent solution and human whole blood passed through High Flux (PES 130) filter were measured in the presence and absence of ultrasound irradiation. Samples were taken from the outlet of the dialyzer every five minutes and the clearance of each constituent was calculated. Results Statistical analysis of the blood equvalent solution and whole blood indicated the clearance of Urea and Inulin in the presence of ultrasound increased (p<0.05), while no significant effects were observed for Creatinine. Conclusions It may be concluded that ultrasound, as a mechanical force, can increase the rate of clearance of some toxins (such as middle and large molecules) in the hemodialysis process.

Authors and Affiliations

M B Shiran, M Barzegar Marvasti, A Shakeri-Zadeh, N Tabkhi, F Farkhondeh, E Kalantar

Keywords

Related Articles

Evaluation of the Effect of Source Geometry on the Output of Miniature X-ray Tube for Electronic Brachytherapy through Simulation

Objective: The use of miniature X-ray source in electronic brachytherapy is on the rise so there is an urgent need to acquire more knowledge on X-ray spectrum production and distribution by a dose. The aim of this resear...

Thyroid Ultrasonography in Differentiation between Graves’ Disease and Hashimoto's Thyroiditis

Purpose: Graves' disease and Hashimoto's thyroiditis are the most common causes of hyper and hypothyroidism, respectively. Differentiation of these 2 diseases, if the patient is euthyroid, may sometimes be extremely diff...

Diagnosis of OCD Patients Using Drawing Features of Bender Gestalt Shapes

Background: Since psychological tests such as questionnaire or drawing tests are almost qualitative, their results carry a degree of uncertainty and sometimes subjectivity. The deficiency of all drawing test is that the...

The Influence of Brass Compensator Thickness and Field Size on Neutron Contamination Spectrum in 18MV Elekta SL 75/25 Medical Linear Accelerator with and without Flattening Filter: A Monte Carlo Study

Background: One of the most significant Intensity Modulated Radiation Therapy treatment benefits is a high target to normal tissue dose ratio. To improve this advantage, an additional accessory such as a compensator is u...

Download PDF file
  • EP ID EP330354
  • DOI -
  • Views 109
  • Downloads 0

How To Cite

M B Shiran, M Barzegar Marvasti, A Shakeri-Zadeh, N Tabkhi, F Farkhondeh, E Kalantar (2017). Enhancement of Toxic Substances Clearance from Blood Equvalent Solution and Human Whole Blood through High Flux Dialyzer by 1 MHz Ultrasound. Journal of Biomedical Physics and Engineering, 7(2), 107-116. https://europub.co.uk/articles/-A-330354