The Influence of Brass Compensator Thickness and Field Size on Neutron Contamination Spectrum in 18MV Elekta SL 75/25 Medical Linear Accelerator with and without Flattening Filter: A Monte Carlo Study

Journal Title: Journal of Biomedical Physics and Engineering - Year 2018, Vol 8, Issue 3

Abstract

Background: One of the most significant Intensity Modulated Radiation Therapy treatment benefits is a high target to normal tissue dose ratio. To improve this advantage, an additional accessory such as a compensator is used to deliver doses. Compensator-based IMRT treatment is usually operated with an energy higher than 10 MV. Photoneutrons, which have high linear energy transfer and radiobiological effectiveness, are produced by colliding high-energy photon beams with linear accelerator structures, then they deliver the unwanted doses to patients and staff. Therefore, the neutron energy spectra should be determined in order to calculate and reduce the photoneutron risk. Objective: We have conducted a comprehensive and precise study on the influence of brass compensator thickness and field size on neutron contamination spectrum in an Elekta SL 75/25 medical linear accelerator with and without the flattening filter by Monte Carlo method. Material and Methods: MCNPX MC Code version 2.6.0 was utilized to simulate the detailed geometry of Elekta SL 75/25 head components based on Linac’s manual. This code includes an important feature to simulate the photo-neutron interactions. Photoneutrons spectrum was calculated after the Linac output benchmarking based on tuning the primary electron beam. Results and Conclusion: Based on the Friedman and Wilcoxon nonparametric tests results (P<0.05), photoneutron fluence directly depends on the field size and compensator thickness. Moreover, the unflattened beam provides lower photoneutron fluence than the flattened beam. Photoneutrons fluence is not negligible in compensator-based IMRT treatment. However, in order to optimize treatment plans, this additional and unwanted dose must be accounted for patientss.

Authors and Affiliations

A. S. Talebi, M. maleki, P. Hejazi, R. Ghorbani

Keywords

Related Articles

New Pseudo-CT Generation Approach from Magnetic Resonance Imaging using a Local Texture Descriptor

Background: One of the challenges of PET/MRI combined systems is to derive an attenuation map to correct the PET image. For that, the pseudo-CT image could be used to correct the attenuation. Until now, most existing sci...

Dosimetry of Critical Organs in Maxillofacial Imaging with Cone-beam Computed Tomography

Background: While the benefits of cone-beam computed tomography (CBCT) are well known in maxillofacial imaging, the use of this modality is not risk-free. Objective: The aim of this study was to evaluate the exposure dos...

Mandibular Trabecular Bone Analysis Using Local Binary Pattern for Osteoporosis Diagnosis

Background: Osteoporosis is a systemic skeletal disease characterized by low bone mineral density (BMD) and micro-architectural deterioration of bone tissue, leading to bone fragility and increased fracture risk. Since P...

Electromagnetic Fields of Mobile Phone Jammer Exposure on Blood Factors in Rats

Background: The increasing demand for using mobile phones has led to increasing mobile phone jammers as well. On the other hand, reports show that exposure to electromagnetic field causes an increase in the incidence of...

The Effect of Long-Term X-Ray Exposure on Human Lymphocyte

The aim of the paper is to investigate effects of long term x-ray exposure on the human lymphocyte, reactive lymphocyte parameters and morphology of lymphocytes in x-ray technicians at Kirkuk hospitals. The study include...

Download PDF file
  • EP ID EP455522
  • DOI -
  • Views 89
  • Downloads 0

How To Cite

A. S. Talebi, M. maleki, P. Hejazi, R. Ghorbani (2018). The Influence of Brass Compensator Thickness and Field Size on Neutron Contamination Spectrum in 18MV Elekta SL 75/25 Medical Linear Accelerator with and without Flattening Filter: A Monte Carlo Study. Journal of Biomedical Physics and Engineering, 8(3), 231-240. https://europub.co.uk/articles/-A-455522