DEVELOPMENT AND VALIDATION OF STABILITY INDICATING RP-HPLC METHOD FOR SIMULTANEOUS ESTIMATION OF ETHINYL ESTRADIOL AND LEVONORGESTREL

Journal Title: Int J of Pharm Res & Analy - Year 2014, Vol 4, Issue 3

Abstract

 ABSTRACT A simple, accurate, precise, sensitive, specific and reliable stability indicating RP-HPLC method was developed for simultaneous estimation of Ethinyl estradiol (EE) and Levonorgestrel (LEV) in Pharmaceutical dosage form. The developed method with mobile phase Acetonitrile: Water (75: 25), Analytica brownee C-18 (150×4.6 mm, 3μm particle size) as a stationary phase and flow rate was 0.8 ml/min. Detection was carried out at 230 nm in PDA detector. The calibration curve of Ethinyl estradiol and Levonorgestrel was found to be linear in the range of 4-14 μg/ml and 20-70 μg/ml respectively. The proposed method has been validated for precision, accuracy, robustness. As the proposed method can effectively separate the drugs from all their degradation products, it can be employed as stability indicating method.

Authors and Affiliations

Tejal Rathod , Anandkumari , D. Captain

Keywords

Related Articles

ATTENUATED TOTAL REFLECTANCE SPECTROSCOPY: AN OVERVIEW

Attenuated Total Reflectance (ATR) is today the most widely used FTIR sampling tool. ATR generally allows qualitative or quantitative analysis of samples with little or no sample preparation which greatly speeds sample a...

 AN OVERVIEW ON HIGH PERFORMANCE CAPILLARY ELECTROPHORESIS

 ABSTRACT High performance Capillary electrophoresis (HPCE) is a separation technique in which the analytes are separated on the basis of differences in their charge-to-size ratios. This makes HPCE especially suit...

 SPECTROPHOTOMETRIC ESTIMATION OF DRUGS USING N-BROMO SUCCINAMIDE AND AMARNTH DYE COUPLE

 Simple, specific, accurate and precise UV–visible spectrophotometric methods have been developed for the estimation of five drugs viz., Dobutamine hydrochloride, Duloxetine hydrochloride, Ondansetron hydrochloride,...

 RP-HPLC METHOD DEVELOPMENT AND VALIDATION FOR SIMULTANEOUS ESTIMATION OF AMLODIPINE BESYLATE AND TELMISARTAN IN TABLET DOSAGE FORM

 A simple, sensitive and rapid reverse phase HPLC method was developed for the simultaneous estimation of Amlodipine besylate and Telmisartan. A Phenominex-luna C18 column (250x4.6 mm i.d 5μ) was used with a mobile...

 DEVELOPMENT AND VALIDATION OF NEW RP-HPLC METHOD FOR PACLITAXEL IN BULK AND ITS PHARMACEUTICAL DOSAGE FORM

 A simple, precise, rapid and accurate reverse phase HPLC method has been developed for the determination of Paclitaxel in bulk and its pharmaceutical dosage form. An enable C18G, 250mm, X4.6mm i.d, 5μm particle siz...

Download PDF file
  • EP ID EP142282
  • DOI -
  • Views 104
  • Downloads 0

How To Cite

Tejal Rathod, Anandkumari, D. Captain (2014).  DEVELOPMENT AND VALIDATION OF STABILITY INDICATING RP-HPLC METHOD FOR SIMULTANEOUS ESTIMATION OF ETHINYL ESTRADIOL AND LEVONORGESTREL. Int J of Pharm Res & Analy, 4(3), 178-185. https://europub.co.uk/articles/-A-142282